期刊论文详细信息
SURFACE & COATINGS TECHNOLOGY 卷:377
Development of multilayer Hydroxyapatite - Ag/TiN-Ti coatings deposited by radio frequency magnetron sputtering with potential application, in the biomedical field
Article; Proceedings Paper
Lenis, J. A.1  Bejarano, G.1  Rico, P.2,3  Gomez Ribelles, J. L.2,3  Bolivar, F. J.1 
[1] Univ Antioquia, Fac Ingn, Ctr Invest Innovat & Desarrollo Mat CIDEMAT, Medellin, Colombia
[2] Univ Politecn Valencia, CBIT, Ctr Biomat & Tissue Engn, Valencia, Spain
[3] Biomed Res Networking Ctr Bioengn Biomat & Nanome, Valencia, Spain
关键词: Magnetron sputtering;    Hydroxyapatite;    Calcium/phosphate ratio;    Microstructure;    Multi-layer coating;    Intermediate layers;    Critical load;    Cytotoxicity;   
DOI  :  10.1016/j.surfcoat.2019.06.097
来源: Elsevier
PDF
【 摘 要 】

The use of composite coatings is emerging as a great alternative to conventional coatings, allowing the combination of different superficial properties that are widely desired in surgical implants, such as osteointegration and bactericidal character, and cannot be provided by one material alone. In the present investigation the effect of the incorporation of a TiN-Ti intermediate bilayer on the chemical composition, structure, morphology, roughness, residual stresses and adhesion of a multi-layer Hydroxyapatite (HA)-Ag coating deposited on Ti-6Al-4V by magnetron sputtering was evaluated. Additionally, the cytotoxicity of the developed system was evaluated by in vitro tests. According to the results obtained, a decrease in the Ca/P ratio from 1.85 to 1.74 was obtained through the deposition of an HA-Ag system on the intermediate bilayer, and the crystallinity of the developed coating was favored. The multi-layer structure was effectively observed by field emission scanning electron microscopy, where it was possible to identify each of the HA, Ag, TiN and Ti layers. Meanwhile, an increase of 7% in crystallite size, a decrease of 36% in residual stresses and an increase of 32% in adhesion were registered for this composite coating compared to the free intermediate bilayer system. Finally, biological evaluation allowed the non-cytotoxic character of the deposited coatings to be confirmed.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_surfcoat_2019_06_097.pdf 3638KB PDF download
  文献评价指标  
  下载次数:5次 浏览次数:0次