| TALANTA | 卷:133 |
| Automated procedure for determination of ammonia in concrete with headspace single-drop micro-extraction by stepwise injection spectrophotometric analysis | |
| Article | |
| Timofeeva, Irina1  Khubaibullin, Ilnur1  Kamencev, Mihail1  Moskvin, Aleksey1  Bulatov, Andrey1  | |
| [1] St Petersburg State Univ, Inst Chem, St Petersburg 198504, Russia | |
| 关键词: Headspace single-drop micro-extraction; Stepwise injection analysis; Ammonia; Concrete; | |
| DOI : 10.1016/j.talanta.2014.04.081 | |
| 来源: Elsevier | |
PDF
|
|
【 摘 要 】
A novel automatic stepwise injection headspace single-drop micro-extraction system is proposed as a versatile approach for automated determination of volatile compounds. The system application is demonstrated for ammonia determination in concrete samples. An ammonia gas was produced from ammonium ions and extracted on-line into 5 mu L, 0.1 M H3PO4 to eliminate the interference effect of concrete species on the ammonia stepwise injection spectrophotometric determination. The linear range was 0.1-1 mg kg(-1) with LOD 30 mu g kg(-1). The sample throughput was 4 h(-1). This system has been successfully applied for the determination of ammonia in concretes. (C) 2014 Elsevier B.V. All rights reserved.
【 授权许可】
Free
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 10_1016_j_talanta_2014_04_081.pdf | 475KB |
PDF