TALANTA | 卷:66 |
Photometric detection in flow analysis systems using integrated PEDDs | |
Article | |
O'Toole, M ; Lau, KT ; Diamond, D | |
关键词: LEDs; optical sensing; pH sensing; transmission measurements; colorimetric analysis; | |
DOI : 10.1016/j.talanta.2005.01.054 | |
来源: Elsevier | |
【 摘 要 】
A novel inexpensive optical-sensing technique has been developed for colorimetric flow analysis. This sensing system employs two LEDs whereby one is used as the light source and the other as a light detector. The LED used as light detector is reverse biased with a 5-V supply so that the photocurrent generated by the incident light discharges the capacitance. Direct digital output is provided by a simple timer circuit that measures the time taken for this discharge process from 5 V (logic 1) to 1.7 V (logic 0). This sensing concept has been applied in flow analysis by constructing an optical flow cell with a pair of LEDs. Calibration of the integrated optical flow cell using a dye resulted in a linear response that obeys the Beer-Lambert law. The flow rate, dynamic range, sensitivity and limits of detection were investigated. The system was also used for pH determination in the range of pH 2.5-6.8 using bromocresol green (BCG). The pK(a) of BCG was successfully determined by this technique. (c) 2005 Elsevier B.V. All rights reserved.
【 授权许可】
Free
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
10_1016_j_talanta_2005_01_054.pdf | 219KB | download |