期刊论文详细信息
TALANTA 卷:66
Photometric detection in flow analysis systems using integrated PEDDs
Article
O'Toole, M ; Lau, KT ; Diamond, D
关键词: LEDs;    optical sensing;    pH sensing;    transmission measurements;    colorimetric analysis;   
DOI  :  10.1016/j.talanta.2005.01.054
来源: Elsevier
PDF
【 摘 要 】

A novel inexpensive optical-sensing technique has been developed for colorimetric flow analysis. This sensing system employs two LEDs whereby one is used as the light source and the other as a light detector. The LED used as light detector is reverse biased with a 5-V supply so that the photocurrent generated by the incident light discharges the capacitance. Direct digital output is provided by a simple timer circuit that measures the time taken for this discharge process from 5 V (logic 1) to 1.7 V (logic 0). This sensing concept has been applied in flow analysis by constructing an optical flow cell with a pair of LEDs. Calibration of the integrated optical flow cell using a dye resulted in a linear response that obeys the Beer-Lambert law. The flow rate, dynamic range, sensitivity and limits of detection were investigated. The system was also used for pH determination in the range of pH 2.5-6.8 using bromocresol green (BCG). The pK(a) of BCG was successfully determined by this technique. (c) 2005 Elsevier B.V. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_talanta_2005_01_054.pdf 219KB PDF download