TALANTA | 卷:99 |
Fast simultaneous determination of BHA and TBHQ antioxidants in biodiesel by batch injection analysis using pulsed-amperometric detection | |
Article | |
Tormin, Thiago F.1  Cunha, Rafael R.1  Richter, Eduardo M.1  Munoz, Rodrigo A. A.1  | |
[1] Univ Fed Uberlandia, Inst Quim, BR-2121 Uberlandia, MG, Brazil | |
关键词: Batch-injection analysis; Multiple-pulse amperometry; Biodiesel; Antioxidants; Simultaneous determination; | |
DOI : 10.1016/j.talanta.2012.06.024 | |
来源: Elsevier | |
【 摘 要 】
We report the first amperometric method for the simultaneous determination of butylated hydroxyanisole (BHA) and tert-butylhydroquinone (TBHQ) in biodiesel using batch-injection analysis (BIA) coupled to pulsed-amperometry. A sequence of potential pulses was selected in order to detect TBHQ and BHA separately in a single injection step at a glassy-carbon electrode. Samples were diluted in 50% v/v hydroethanolic solution with 0.1 mol L-1 HClO4 (supporting electrolyte) before injection using an electronic pipette. The method is highly precise (RSD < 1%, n=10), fast (170 injections h(-1)), accurate (recovery between 100 and 110%), presents low detection limits (73 and 75 nmol L-1 BHA and TBHQ respectively), and can be easily adapted for on-site determinations. (C) 2012 Elsevier B.V. All rights reserved.
【 授权许可】
Free
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
10_1016_j_talanta_2012_06_024.pdf | 388KB | download |