In this work, the hexadentate 3,4-hydroxypyridinone ligand was used as reagent for the spectrophotometric quantification of iron(III) in fresh and sea waters, using a micro sequential injection lab-on-valve (mu SI-LOV) system in a solid phase spectrometry (SPS) mode. To implement SPS, thus eliminating the sample matrix, a packed column in the flow cell was used; the chosen sorbent was Nitrilotriacetic Acid Supefflow resin (NTA). The possibility of performing an analytical curve resorting to just one standard was also demonstrated. The consumption of the hexadentante ligand was about 3014 per determination and the effluent production lower than 2.5 mL. The dynamic concentration range was 0.45-9.0 mu mol L-1, with a limit of detection of 0.13 mu mol L-1 and limit of quantification 0.43 mu mol L-1. The proposed iSI-LOV-SPS methodology was successfully applied to river, ground, estuarine, tap, and sea waters.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_talanta_2018_08_063.pdf 431KB PDF download
  文献评价指标  
  下载次数:0次 浏览次数:0次