Detection limits of chemical sensors: Applications and misapplications
Article
Loock, Hans-Peter1  Wentzell, Peter D.2 
[1] Queens Univ, Dept Chem, Kingston, ON K7L 3N6, Canada
[2] Dalhousie Univ, Dept Chem, Halifax, NS B3H 4J3, Canada
关键词: Limit of detection;    Sensitivity;    Calibration curve;    Confidence interval;    Fiber optic sensors;    Photoacoustic sensor;    Absorption sensor;    Refractive index sensor;   
DOI  :  10.1016/j.snb.2012.06.071
来源: Elsevier
PDF
【 摘 要 】

The limit of detection (LOD) and the sensitivity of a chemical sensor are defined using IUPAC guidelines. The LOD from simulated and experimental data is calculated from a calibration curve using a simple statistical model that was implemented into a spreadsheet program. This definition of the LOD is compared with the commonly used definition of the LOD, which is based on the product of sensitivity and the theoretical instrument resolution. Crown Copyright (C) 2012 Published by Elsevier B.V. All rights reserved.