期刊论文详细信息
SENSORS AND ACTUATORS B-CHEMICAL 卷:199
New characterization methods for monitoring small resonant frequency variation: Experimental tests in the case of hydrogen detection with uncoated silicon microcantilever-based sensors
Article
Boudjiet, M. T.1  Bertrand, J.2  Pellet, C.1  Dufour, I.1 
[1] Univ Bordeaux, IMS, UMR 5218, F-33400 Talence, France
[2] Andra, F-92298 Chatenay Malabry, France
关键词: Resonant frequency;    Quality factor;    Uncoated silicon microcantilever;    Chemical detection;    Signal-to-noise ratio;   
DOI  :  10.1016/j.snb.2014.03.098
来源: Elsevier
PDF
【 摘 要 】

The uncoated silicon microcantilever (USMC) operated in the dynamic mode is a new concept in the field of microcantilever-based chemical sensors. Due to the absence of a sensitive layer, this kind of microsensor can only be used for specific applications where it is known that only one chemical species may be varying in concentration, such as monitoring hydrogen release in radioactive waste disposal facilities. Usually, the relative variation of the USMC resonant frequency expected for low concentrations (<= 2%) of hydrogen in nitrogen is below 50 ppm. As a result, the measurement of both the resonant frequencies, f(r), and the quality factor, Q, by classical methods, based on the gain spectrum (resonant peak and 3 dB bandwidth), is not sufficiently accurate. In this paper, new measurement methods for monitoring fr and Q variations are proposed: (1) variation of gain and phase at fixed frequencies and (2) polynomial approximations of gain and phase spectra. The performance study of these characterization methods shows that monitoring f(r) by using phase linearization yields the best signal-to-noise ratio (e.g., 100 at 0.6% of H-2 in N-2), with 0.02% as a limit of detection for hydrogen. (C) 2014 Elsevier B.V. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_snb_2014_03_098.pdf 2029KB PDF download