Graphene-polyaniline composite film modified electrode for voltammetric determination of 4-aminophenol
Article
Fan, Yang1  Liu, Jin-Hang1  Yang, Chun-Peng2  Yu, Meng1  Liu, Peng1 
[1] Xinyang Normal Univ, Coll Chem & Chem Engn, Xinyang 464000, Peoples R China
[2] Chinese Acad Sci, Qingdao Inst Bioenergy & Bioproc Technol, Qingdao 266101, Peoples R China
关键词: Graphene;    Polyaniline;    Nanocomposite;    Aminophenol;    Electrochemical sensor;   
DOI  :  10.1016/j.snb.2011.05.053
来源: Elsevier
PDF
【 摘 要 】

An electrochemical sensor based on graphene-polyaniline (GR-PANI) nanocomposite for voltammetric determination of 4-aminophenol (4-AP) is presented. The electrochemical behavior of 4-AP at the GR-PANI composite film modified glassy carbon electrode (GCE) was investigated by cyclic voltammetry. 4-AP exhibits enhanced voltammetric response at GR-PANI modified GCE. This electrochemical sensor shows a favorable analytical performance for 4-AP detection with a detection limit of 6.5 x 10(-8) M and high sensitivity of 604.2 mu A mM(-1). Moreover. 4-AP and paracetamol can be detected simultaneously without interference of each other in a large dynamic range. (C) 2011 Elsevier B.V. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_snb_2011_05_053.pdf 644KB PDF download