Malaria Journal | |
Evaluation of the rapid diagnostic test CareStart pLDH Malaria (Pf-pLDH/pan-pLDH) for the diagnosis of malaria in a reference setting | |
Research | |
Jan Jacobs1  Jessica Maltha1  Philippe Gillet1  Lieselotte Cnops1  Emmanuel Bottieau1  Annelies Scheirlinck1  Marjan Van Esbroeck1  Marloes Heutmekers2  | |
[1] Department of Clinical Sciences, Institute of Tropical Medicine, Antwerp, Belgium;Faculty of Health, Medicine and Life Sciences (FHML), Maastricht University, Maastricht, The Netherlands; | |
关键词: Malaria; Plasmodium; Diagnostic; Evaluation; RDT; Rapid diagnostic tests; pLDH; Lactate dehydrogenase; CareStart; | |
DOI : 10.1186/1475-2875-11-204 | |
received in 2012-02-21, accepted in 2012-06-03, 发布年份 2012 | |
来源: Springer | |
【 摘 要 】
BackgroundThe present study evaluated CareStart pLDH Malaria, a three-band rapid diagnostic test detecting Plasmodium falciparum- specific parasite lactate dehydrogenase (Pf-pLDH) and pan Plasmodium-specific pLDH (pan-pLDH) in a reference setting.MethodsCareStart pLDH was retrospectively and prospectively assessed with a panel of stored (n = 498) and fresh (n = 77) blood samples obtained in international travelers suspected of malaria. Both panels comprised all four Plasmodium species; the retrospective panel comprised also Plasmodium negative samples. The reference method was microscopy corrected by PCR. The prospective panel was run side-to-side with OptiMAL (Pf-pLDH/pan-pLDH) and SDFK60 (histidine-rich protein-2 (HRP-2)/pan-pLDH).ResultsIn the retrospective evaluation, overall sensitivity for P. falciparum samples (n = 247) was 94.7%, reaching 98.7% for parasite densities > 1,000/μl. Most false negative results occurred among samples with pure gametocytaemia (2/12, 16.7%) and at parasite densities ≤ 100/μl (7/12, 58.3%). None of the Plasmodium negative samples (n = 96) showed visible test lines. Sensitivities for Plasmodium vivax (n = 70), Plasmodium ovale (n = 69) and Plasmodium malariae (n = 16) were 74.3%, 31.9% and 25.0% respectively. Wrong species identification occurred in 10 (2.5%) samples and was mainly due to P. vivax samples reacting with the Pf-pLDH test line. Overall, Pf-pLDH test lines showed higher line intensities compared to the pan-pLDH lines (67.9% and 23.0% medium and strong line intensities for P. falciparum). In the prospective panel (77 Plasmodium-positive samples), CareStart pLDH showed higher sensitivities for P. falciparum compared to OptiMAL (p = 0.008), lower sensitivities for P. falciparum as compare to SDFK60 (although not reaching statistical significance, p = 0.08) and higher sensitivities for P. ovale c ompared to both OptiMAL (p = 0.03) and SDFK60 (p = 0.01). Inter-observer and test reproducibility were good to excellent.ConclusionCareStart pLDH performed excellent for the detection of P. falciparum, well for P. vivax, but poor for P. ovale and P. malariae.
【 授权许可】
Unknown
© Heutmekers et al.; licensee BioMed Central Ltd. 2012. This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202311109606563ZK.pdf | 422KB | download |
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]