期刊论文详细信息
Journal of Cardiovascular Magnetic Resonance
Safe performance of magnetic resonance of the heart in patients with magnetic resonance conditional pacemaker systems: the safety issue of the ESTIMATE study
Research
Karin Thudt1  Christian G Wollmann1  Harald Mayr1  Sebastian Globits1  Erich Salomonowitz2  Bernd Kaiser2 
[1] Department of Cardiology, Hospital of St. Pölten-Lilienfeld, St. Pölten, Austria;Karl Landsteiner Society, Institute of Cardiovascular Research, St. Pölten, Austria;Department of Radiology and Interventional Angiology, Hospital of St. Pölten-Lilienfeld, St. Pölten, Austria;
关键词: Permanent pacemaker;    MR conditional;    Magnet resonance;    Heart MR;    Complications;   
DOI  :  10.1186/1532-429X-16-30
 received in 2014-01-02, accepted in 2014-03-28,  发布年份 2014
来源: Springer
PDF
【 摘 要 】

BackgroundNo published data exist about the safety of diagnostic magnetic resonance (MR) of the heart performed in a larger series of patients implanted with MR conditional pacemakers (PM). The purpose of our study is to analyse safety and potential alterations of electrical lead parameters in patients implanted with the EnRhythm/Advisa MRI SureScan PM with 5086MRI leads (Medtronic Inc.) during and after MR of the heart at 1.5 Tesla.MethodsPatients enrolled in this single center pilot study who underwent non-clinically indicated diagnostic MR of the heart were included in this analysis. Heart MR was performed for analyses of potential changes in right and left ventricular functional parameters under right ventricular pacing at 80 and 110 bpm. Atrial/ventricular sensing, atrial/ventricular pacing capture threshold [PCT], and pacing impedances were assessed immediately before, during, and immediately after MR, as well at 3 and 15 months post MR.ResultsThirty-six patients (mean age 69 ± 13 years; high degree AV block 18 [50%]) underwent MR of the heart. No MR related adverse events occurred during MR or thereafter. Ventricular sensing differed significantly between the FU immediately after MR (10.3 ± 5.3 mV) and the baseline FU (9.8 ± 5.3 mV; p < 0.05). Despite PCT [V/0.4ms] was not significantly different between the FUs (baseline: 0.84 ± 0.27; in-between MR scans: 0.82 ± 0.27; immediately after MR: 0.84 ± 0.24; 3-month: 0.85 ± 0.23; 15-month: 0.90 ± 0.67; p = ns), 7 patients (19%) showed PCT increases by 100% (max. PCT measured: 1.0 V) at the 3-month FU compared to baseline. RV pacing impedance [Ω/5V] differed significantly at the FU in-between MR scans (516 ± 47), and at the 15-month FU (482 ± 58) compared to baseline (508 ± 75).ConclusionThe results of our study suggest MR of the heart to be safe in patients with the MR conditional EnRhythm/Advisa system, albeit although noticeable but clinically irrelevant ventricular PCT changes were observed.

【 授权许可】

CC BY   
© Wollmann et al.; licensee BioMed Central Ltd. 2014

【 预 览 】
附件列表
Files Size Format View
RO202311106577054ZK.pdf 382KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  文献评价指标  
  下载次数:3次 浏览次数:1次