Journal of Cardiovascular Magnetic Resonance
Towards real-time cardiovascular magnetic resonance guided transarterial CoreValve implantation: in vivo evaluation in swine
Research
Raimund Erbel1  Holger Eggebrecht1  Philipp Kahlert1  Björn Plicht1  Nina Parohl2  Mark E Ladd2  Juliane Albert2  Lena Schäfer2  Harald H Quick3  Renate Reinhardt4  Gernot M Kaiser4  Ian McDougall5  Brad Decker5 
[1] Department of Cardiology, West-German Heart Center Essen, Essen University Hospital, University Duisburg-Essen, Hufelandstrasse 55, 45122, Essen, Germany;Department of Diagnostic and Interventional Radiology and Neuroradiology, Essen University Hospital, University of Duisburg-Essen, Hufelandstrasse 55, 45122, Essen, Germany;Department of Diagnostic and Interventional Radiology and Neuroradiology, Essen University Hospital, University of Duisburg-Essen, Hufelandstrasse 55, 45122, Essen, Germany;Institute of Medical Physics, Friedrich-Alexander-University Erlangen-Nürnberg, Henkestrasse 91, 91052, Erlangen, Germany;Department of General, Visceral and Transplantation Surgery, Essen University Hospital, University Duisburg-Essen, Hufelandstrasse 55, 45122, Essen, Germany;Evasc Medical Systems, 107-1099 West 8th Avenue, V6H 1C3, Vancouver, BC, Canada;
关键词: Aortic stenosis;    Transcatheter aortic valve implantation;    Cardiovascular magnetic resonance;    Real-time;   
DOI  :  10.1186/1532-429X-14-21
 received in 2011-10-14, accepted in 2012-03-27,  发布年份 2012
来源: Springer
PDF
【 摘 要 】

BackgroundReal-time cardiovascular magnetic resonance (rtCMR) is considered attractive for guiding TAVI. Owing to an unlimited scan plane orientation and an unsurpassed soft-tissue contrast with simultaneous device visualization, rtCMR is presumed to allow safe device navigation and to offer optimal orientation for precise axial positioning. We sought to evaluate the preclinical feasibility of rtCMR-guided transarterial aortic valve implatation (TAVI) using the nitinol-based Medtronic CoreValve bioprosthesis.MethodsrtCMR-guided transfemoral (n = 2) and transsubclavian (n = 6) TAVI was performed in 8 swine using the original CoreValve prosthesis and a modified, CMR-compatible delivery catheter without ferromagnetic components.ResultsrtCMR using TrueFISP sequences provided reliable imaging guidance during TAVI, which was successful in 6 swine. One transfemoral attempt failed due to unsuccessful aortic arch passage and one pericardial tamponade with subsequent death occurred as a result of ventricular perforation by the device tip due to an operating error, this complication being detected without delay by rtCMR. rtCMR allowed for a detailed, simultaneous visualization of the delivery system with the mounted stent-valve and the surrounding anatomy, resulting in improved visualization during navigation through the vasculature, passage of the aortic valve, and during placement and deployment of the stent-valve. Post-interventional success could be confirmed using ECG-triggered time-resolved cine-TrueFISP and flow-sensitive phase-contrast sequences. Intended valve position was confirmed by ex-vivo histology.ConclusionsOur study shows that rtCMR-guided TAVI using the commercial CoreValve prosthesis in conjunction with a modified delivery system is feasible in swine, allowing improved procedural guidance including immediate detection of complications and direct functional assessment with reduction of radiation and omission of contrast media.

【 授权许可】

CC BY   
© Kahlert et al; licensee BioMed Central Ltd. 2012

【 预 览 】
附件列表
Files Size Format View
RO202311105740410ZK.pdf 4865KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  文献评价指标  
  下载次数:5次 浏览次数:1次