期刊论文详细信息
BMC Medical Imaging
Registration error of the liver CT using deformable image registration of MIM Maestro and Velocity AI
Research Article
Kazunori Nitta1  Kayoko Ohnishi2  Yoshiko Oshiro2  Hideyuki Sakurai2  Hitoshi Ishikawa2  Koji Tsuboi2  Masashi Mizumoto2  Teruhito Aihara2  Haruko Numajiri2  Toshiyuki Okumura2  Toshiyuki Terunuma2  Nobuyoshi Fukumitsu2 
[1] Division of Radiology, Ibaraki Prefectural Central Hospital, 6528, Koibuchi, Kasama, Japan;Proton Medical Research Center, University of Tsukuba, 1-1-1, Tennoudai, Tsukuba, Japan;
关键词: Deformable image registration;    Rigid image registration;    Liver;    Proton beam therapy;   
DOI  :  10.1186/s12880-017-0202-z
 received in 2016-12-09, accepted in 2017-04-20,  发布年份 2017
来源: Springer
PDF
【 摘 要 】

BackgroundUnderstanding the irradiated area and dose correctly is important for the reirradiation of organs that deform after irradiation, such as the liver. We investigated the spatial registration error using the deformable image registration (DIR) software products MIM Maestro (MIM) and Velocity AI (Velocity).MethodsImage registration of pretreatment computed tomography (CT) and posttreatment CT was performed in 24 patients with liver tumors. All the patients received proton beam therapy, and the follow-up period was 4–14 (median: 10) months. We performed DIR of the pretreatment CT and compared it with that of the posttreatment CT by calculating the dislocation of metallic markers (implanted close to the tumors).ResultsThe fiducial registration error was comparable in both products: 0.4–32.9 (9.3 ± 9.9) mm for MIM and 0.5–38.6 (11.0 ± 10.0) mm for Velocity, and correlated with the tumor diameter for MIM (r = 0.69, P = 0.002) and for Velocity (r = 0.68, P = 0.0003). Regarding the enhancement effect, the fiducial registration error was 1.0–24.9 (7.4 ± 7.7) mm for MIM and 0.3–29.6 (8.9 ± 7.2) mm for Velocity, which is shorter than that of plain CT (P = 0.04, for both).ConclusionsThe DIR performance of both MIM and Velocity is comparable with regard to the liver. The fiducial registration error of DIR depends on the tumor diameter. Furthermore, contrast-enhanced CT improves the accuracy of both MIM and Velocity.Institutional review board approvalH28-102; July 14, 2016 approved.

【 授权许可】

CC BY   
© The Author(s). 2017

【 预 览 】
附件列表
Files Size Format View
RO202311100938918ZK.pdf 5517KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  文献评价指标  
  下载次数:11次 浏览次数:4次