Insights into Imaging | |
Optimal combination periprosthetic vasculature visualization and metal artifact reduction by spectral computed tomography using virtual monoenergetic images in total hip arthroplasty | |
Original Article | |
Qiang Cheng1  Haitao Yang2  Ziyi Tang2  Jie Zhao2  Qiqi Wang2  Yuchan Lv2  Chuan Liu2  Yuxi Luo2  | |
[1] Department of Orthopedics, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China;Department of Radiology, The First Affiliated Hospital of Chongqing Medical University, 1 Youyi Road, Yuzhong District, 400016, Chongqing, China; | |
关键词: Energy spectrum CT; Virtual monoenergetic imaging; Metal artifact reduction; Periprosthetic vasculature; | |
DOI : 10.1186/s13244-023-01533-3 | |
received in 2023-07-23, accepted in 2023-09-29, 发布年份 2023 | |
来源: Springer | |
【 摘 要 】
ObjectivesTo investigate the optimal parameters of spectral CT for preferably visualizing the periprosthetic vasculature and metal artifact reduction (MAR) in total hip arthroplasty (THA).MethodsA total of 34 THA of 30 patients were retrospectively included. Image reconstructions included conventional image (CI), CI combined with MAR (CIMAR), and virtual monoenergetic images (VMI) combined with MAR (VMIMAR) at 50–120 keV. The attenuation and standard deviation of the vessel and artifact, and the width of artifact were measured. Qualitative scoring was evaluated including the vascular contour, the extent of artifact, and overall diagnostic evaluation.ResultsThe attenuation, noise of the vessel and artifact, and the width of artifact decreased as the energy level increased (p < 0.001). The downtrend was relatively flat at 80–120 keV, and the vascular attenuation dropped to 200 HU at 90 keV. The qualitative rating of vascular contour was significantly higher at CIMAR (3.47) and VMIMAR 60–80 keV (2.82–3.65) compared with CI (2.03) (p ≤ 0.029), and the highest score occurred at 70 and 80 keV (3.65 and 3.56). The score of the extent of artifact was higher at VMIMAR 80 keV than CIMAR (3.53 VS 3.12, p = 0.003). The score of the overall diagnostic evaluation was higher at VMIMAR 70 and 80 keV (3.32 and 3.53, respectively) than CIMAR (3.12) (p ≤ 0.035).ConclusionEighty kiloelectron volts on VMIMAR, providing satisfactorily reduced metal artifacts and improved vascular visualization, can be an optimal recommended parameter of spectrum CT for the assessment of periprosthetic vasculature in THA patients.Critical relevance statementThe metal artifact is gradually reducing with increasing energy level; however, the vascular visualization is worsening. The vascular visualization is terrible above 100 keV, while the vessel is disturbed by artifacts below 70 keV. The best performance is found at 80 keV.Key points• VMIMAR can provide both reduced metal artifacts and improved vascular visualization.• Eighty kiloelectron volts on VMIMAR performs best in vascular visualization of total hip arthroplasty patients.• Energy spectrum CT is recommended for routine use in patients with total hip arthroplasty.Graphical Abstract
【 授权许可】
CC BY
© European Society of Radiology (ESR) 2023
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202311100173610ZK.pdf | 2462KB | download | |
Fig. 7 | 684KB | Image | download |
MediaObjects/12951_2023_2144_MOESM1_ESM.docx | 15232KB | Other | download |
12951_2015_155_Article_IEq54.gif | 1KB | Image | download |
12936_2017_1932_Article_IEq37.gif | 1KB | Image | download |
12936_2017_1963_Article_IEq63.gif | 1KB | Image | download |
Fig. 2 | 192KB | Image | download |
【 图 表 】
Fig. 2
12936_2017_1963_Article_IEq63.gif
12936_2017_1932_Article_IEq37.gif
12951_2015_155_Article_IEq54.gif
Fig. 7
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]
- [28]
- [29]
- [30]
- [31]
- [32]
- [33]
- [34]
- [35]
- [36]
- [37]
- [38]
- [39]
- [40]
- [41]
- [42]
- [43]
- [44]