BMC Ophthalmology
Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes
Research Article
Dolores de Fez1  María T. Caballero1  Inmaculada Cabezos1  Vicente J. Camps1  David P. Piñero1  Alberto López-Navarro2 
[1] Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Crta San Vicente del Raspeig s/n 03016, San Vicente del Raspeig, Alicante, Spain;Optometry Clinic, Prevention Service of the University of Alicante, Alicante, Spain;
关键词: Anterior chamber depth;    Pachymetry;    White-to-white;    Irido-corneal angle;    Scheimpflug imaging;    VX120 system;   
DOI  :  10.1186/s12886-017-0589-x
 received in 2016-10-03, accepted in 2017-10-16,  发布年份 2017
来源: Springer
PDF
【 摘 要 】

BackgroundTo evaluate the intrasession repeatability of anterior chamber depth (ACD), central (CCT) and peripheral corneal thickness (PCT), white-to-white diameter (WTW), and irido-corneal angle (IA) measurements obtained with a multidiagnostic device in healthy eyes.MethodsA total of 107 eyes of 107 patients ranging in age from 23 to 65 years were examined with the VX120 system (Visionix-Luneau Technologies). Three consecutive measurements were obtained with this device to assess the intrasession repeatability of ACD, CCT, PCT at different nasal and temporal locations, WTW, and nasal and temporal IA. Data analysis included the calculation of within-subject standard deviation (Sw), intrasubject precision (1.96xSw), coefficient of variation (CV) and intraclass correlation coefficient (ICC).ResultsThe Sw and CV for ACD was 0.03 mm and 1.16%, respectively, with an ICC of 0.992. The Sw values for central and peripheral pachymetric measurements were below 9 μm, with CV of less than 1.6% and ICC of 0.976 or higher. For IA measurements, Sw values of 0.84 or lower were found, with a CV between 1 and 2%, and an ICC of more than 0.970. The Sw for WTW was 0.24 mm and the CV was 1.95%. No statistically significant correlations were found between any anatomical parameter evaluated and their Sw and CV values associated (−0.220 ≤ r ≤ 0.204, p ≥ 0.125).ConclusionsThe VX120 system is able to provide repeatable measurements of anatomical parameters in healthy eyes. Inter-observer repeatability should be evaluated in future studies.

【 授权许可】

CC BY   
© The Author(s). 2017

【 预 览 】
附件列表
Files Size Format View
RO202311099738808ZK.pdf 708KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  文献评价指标  
  下载次数:2次 浏览次数:0次