Assessment of a novel biomechanical fracture model for distal radius fractures
Research Article
Hannes Traxler1  Sebastian Felix Baumbach2  Enrico Dall’Ara3  Philippe K Zysset4  Patrick Weninger5  Anna Antoni5  Martin Dörr6 
[1] Center of Anatomy and Cell Biology,Department of Systematic Anatomy, Medical University Vienna, Vienna, Austria;Department of Surgery, Experimental Surgery and Regenerative Medicine, Ludwig-Maximilians-University Munich, Munich, Germany;Institute of Lightweight Design and Structural Biomechanics, Vienna University of Technology, Vienna, Austria;Institute of Surgical Technology and Biomechanics, University of Bern, Bern, Switzerland;Lorenz Boehler Trauma Hospital, Ludwig Boltzmann Institute for Experimental and Clinical Traumatology, Cluster for Tissue Regeneration, Vienna, Austria;Munich Cancer Registry; IBE / Clinic Großhadern, Ludwig-Maximilians University Munich, Munich, Germany;
关键词: Biomechanical fracture model;    Distal radius fracture;    Fresh-frozen;    Plate osteosynthesis;   
DOI  :  10.1186/1471-2474-13-252
 received in 2012-03-30, accepted in 2012-12-10,  发布年份 2012
来源: Springer
PDF
【 摘 要 】

BackgroundDistal radius fractures (DRF) are one of the most common fractures and often need surgical treatment, which has been validated through biomechanical tests. Currently a number of different fracture models are used, none of which resemble the in vivo fracture location. The aim of the study was to develop a new standardized fracture model for DRF (AO-23.A3) and compare its biomechanical behavior to the current gold standard.MethodsVariable angle locking volar plates (ADAPTIVE, Medartis) were mounted on 10 pairs of fresh-frozen radii. The osteotomy location was alternated within each pair (New: 10 mm wedge 8 mm / 12 mm proximal to the dorsal / volar apex of the articular surface; Gold standard: 10 mm wedge 20 mm proximal to the articular surface). Each specimen was tested in cyclic axial compression (increasing load by 100 N per cycle) until failure or −3 mm displacement. Parameters assessed were stiffness, displacement and dissipated work calculated for each cycle and ultimate load. Significance was tested using a linear mixed model and Wald test as well as t-tests.Results7 female and 3 male pairs of radii aged 74 ± 9 years were tested. In most cases (7/10), the two groups showed similar mechanical behavior at low loads with increasing differences at increasing loads. Overall the novel fracture model showed a significant different biomechanical behavior than the gold standard model (p < 0,001). The average final loads resisted were significantly lower in the novel model (860 N ± 232 N vs. 1250 N ± 341 N; p = 0.001).ConclusionThe novel biomechanical fracture model for DRF more closely mimics the in vivo fracture site and shows a significantly different biomechanical behavior with increasing loads when compared to the current gold standard.

【 授权许可】

CC BY   
© Baumbach et al.; licensee BioMed Central Ltd. 2012

【 预 览 】
附件列表
Files Size Format View
RO202311099345388ZK.pdf 1301KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  文献评价指标  
  下载次数:1次 浏览次数:1次