期刊论文详细信息
BMC Musculoskeletal Disorders
Biomechanical comparison of a novel transoral atlantoaxial anchored cage with established fixation technique - a finite element analysis
Research Article
Feng Xu1  Hui Kang1  Ran Ding1  Xian-hua Cai2  Bao-cheng Zhang2  Hai-bo Liu3  Zhi-hua Wang3  Xiao-qing Luo4 
[1] Department of Orthopaedics, Wuhan General Hospital of Guangzhou Command of PLA, Wuhan 430070, China;Department of Orthopaedics, Wuhan General Hospital of Guangzhou Command of PLA, Wuhan 430070, China;Southern Medical University, Guangzhou 510515, China;Institute of Applied Mechanics and Biomedical Engineering, Taiyuan University of Technology, Taiyuan 030024, China;The School of Internet of Things, Jiangnan University, Wuxi 214122, China;
关键词: Bone Graft;    Lateral Mass;    Plate Model;    Basilar Invagination;    Cage Subsidence;   
DOI  :  10.1186/s12891-015-0662-7
 received in 2015-03-15, accepted in 2015-08-06,  发布年份 2015
来源: Springer
PDF
【 摘 要 】

BackgroundThe transoral atlantoaxial reduction plate (TARP) fixation has been introduced to achieve reduction, decompression, fixation and fusion of C1–C2 through a transoral-only approach. However, it may also be associated with potential disadvantages, including dysphagia and load shielding of the bone graft. To prevent potential disadvantages related to TARP fixation, a novel transoral atlantoaxial fusion cage with integrated plate (Cage + Plate) device for stabilization of the C1-C2 segment is designed. The aims of the present study were to compare the biomechanical differences between Cage + Plate device and Cage + TARP device for the treatment of basilar invagination (BI) with irreducible atlantoaxial dislocation (IAAD).MethodsA detailed, nonlinear finite element model (FEM) of the intact upper cervical spine had been developed and validated. Then a FEM of an unstable BI model treated with Cage + Plate fixation, was compared to that with Cage + TARP fixation. All models were subjected to vertical load with pure moments in flexion, extension, lateral bending and axial rotation. Range of motion (ROM) of C1-C2 segment and maximum von Mises Stress of the C2 endplate and bone graft were quantified for the two devices.ResultsBoth devices significantly reduced ROM compared with the intact state. In comparison with the Cage + Plate model, the Cage + TARP model reduced the ROM by 82.5 %, 46.2 %, 10.0 % and 74.3 % in flexion, extension, lateral bending, and axial rotation. The Cage + Plate model showed a higher increase stresses on C2 endplate and bone graft than the Cage + TARP model in all motions.ConclusionsOur results indicate that the novel Cage + Plate device may provide lower biomechanical stability than the Cage + TARP device in flexion, extension, and axial rotation, however, it may reduce stress shielding of the bone graft for successful fusion and minimize the risk of postoperative dysphagia. Clinical trials are now required to validate the reproducibility and advantages of our findings using this anchored cage for the treatment of BI with IAAD.

【 授权许可】

CC BY   
© Zhang et al. 2015

【 预 览 】
附件列表
Files Size Format View
RO202311098481174ZK.pdf 3570KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  • [38]
  • [39]
  • [40]
  • [41]
  • [42]
  • [43]
  • [44]
  • [45]
  • [46]
  • [47]
  • [48]
  • [49]
  文献评价指标  
  下载次数:1次 浏览次数:1次