BMC Medical Imaging | |
An open-source software tool for the generation of relaxation time maps in magnetic resonance imaging | |
Software | |
Hassan Abdel-Aty1  Ralf Wassmuth1  Rainer Dietz1  Andre Rudolph1  Jeanette Schulz-Menger1  Daniel R Messroghli2  Titus Kühne3  | |
[1] Cardiac MRI Unit, Franz-Volhard-Klinik, Charité University Medicine, Berlin, Germany;Cardiac MRI Unit, Franz-Volhard-Klinik, Charité University Medicine, Berlin, Germany;Cardiovascular Imaging Unit, Department of Congenital Heart Defects and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Germany;Cardiovascular Imaging Unit, Department of Congenital Heart Defects and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Germany; | |
关键词: Source Image; Magnetic Resonance System; Quantitative Magnetic Resonance; DICOM Viewer; DICOM Header; | |
DOI : 10.1186/1471-2342-10-16 | |
received in 2010-04-01, accepted in 2010-07-30, 发布年份 2010 | |
来源: Springer | |
【 摘 要 】
BackgroundIn magnetic resonance (MR) imaging, T1, T2 and T2* relaxation times represent characteristic tissue properties that can be quantified with the help of specific imaging strategies. While there are basic software tools for specific pulse sequences, until now there is no universal software program available to automate pixel-wise mapping of relaxation times from various types of images or MR systems. Such a software program would allow researchers to test and compare new imaging strategies and thus would significantly facilitate research in the area of quantitative tissue characterization.ResultsAfter defining requirements for a universal MR mapping tool, a software program named MRmap was created using a high-level graphics language. Additional features include a manual registration tool for source images with motion artifacts and a tabular DICOM viewer to examine pulse sequence parameters. MRmap was successfully tested on three different computer platforms with image data from three different MR system manufacturers and five different sorts of pulse sequences: multi-image inversion recovery T1; Look-Locker/TOMROP T1; modified Look-Locker (MOLLI) T1; single-echo T2/T2*; and multi-echo T2/T2*. Computing times varied between 2 and 113 seconds. Estimates of relaxation times compared favorably to those obtained from non-automated curve fitting. Completed maps were exported in DICOM format and could be read in standard software packages used for analysis of clinical and research MR data.ConclusionsMRmap is a flexible cross-platform research tool that enables accurate mapping of relaxation times from various pulse sequences. The software allows researchers to optimize quantitative MR strategies in a manufacturer-independent fashion. The program and its source code were made available as open-source software on the internet.
【 授权许可】
CC BY
© Messroghli et al; licensee BioMed Central Ltd. 2010
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202311095654310ZK.pdf | 4653KB | download |
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]