| Results of the cementless Plasmacup in revision total hip arthroplasty: a retrospective study of 72 cases with an average follow-up of eight years | |
| Research Article | |
| Stefan Lakemeier1  Guenter Aurand1  Thomas J Heyse1  Susanne Fuchs-Winkelmann1  Markus D Schofer1  Nina Timmesfeld2  | |
| [1] Department of Orthopaedics and Rheumathology, Baldingerstraße, 35043 Marburg, University Hospital Gießen and Marburg, location Marburg, Germany;Institute of Medical Biometry and Epidemiology, Bunsenstraße 3, 35037 Marburg, Phillips-University Marburg, Germany; | |
| 关键词: Aseptic Loosening; Radiolucent Line; Acetabular Defect; Paprosky Type; Acetabular Bone Stock; | |
| DOI : 10.1186/1471-2474-11-101 | |
| received in 2009-08-29, accepted in 2010-05-27, 发布年份 2010 | |
| 来源: Springer | |
PDF
|
|
【 摘 要 】
BackgroundThere are multiple revision implant systems currently available for socket revision in revision total hip arthroplasty. Up until now, not all of these systems have been followed up with regards to their long-term use as a revision implantation.For the first time, this study presents the hemispherical porous-coated socket Plasmacup SC, produced by Aesculap, Tuttlingen, Germany, and the clinical and radiological mid-term results of this revision cup implant.MethodsOver a period of ten years the Plasmacup SC press-fit-cup was used as a revision implant in 72 consecutive aseptic cases which were included in this retrospective study. The mean follow-up period was 8 years. Bone graft transplantation was performed in 32% of all cases. In 90%, the cup was fixed with additional screws. The follow-up radiographs were analysed with regards to cup migration, osteointegration and osteolysis in the DeLee zones using a computer aided program taking the teardrop figure as a main point of reference. For clinical evaluation the Harris-Hip-Score and the WOMAC-Score were utilized.ResultsAt the follow up examination, the mean Harris-Hip-Score was 83.5 points and the mean WOMAC-Score 34.7 points. 93% of all patients were satisfied with the result of the operation. No aseptic cup loosening could be observed and only one cup had to be removed due to infection. No significant longitudinal or transversal cup migration could be observed.ConclusionAesculap's Plasmacup SC is suitable as a cementless cup revision implant. There is stable cup osteointegration, post press-fit implantation, even in the case of major acetabular bone defects.
【 授权许可】
CC BY
© Lakemeier et al; licensee BioMed Central Ltd. 2010
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202311095566035ZK.pdf | 801KB |
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]
- [28]
- [29]
- [30]
- [31]
- [32]
- [33]
- [34]
- [35]
- [36]
- [37]
- [38]
- [39]
- [40]
- [41]
PDF