BMC Ophthalmology | |
In vitro comparative optical bench analysis of a spherical and aspheric optic design of the same IOL model | |
Research Article | |
Gerd U. Auffarth1  Ramin Khoramnia1  Tamer Tandogan1  Stephanie Liebing1  Chul Y. Choi2  Christian Mayer3  | |
[1] David J Apple International Laboratory for Ocular Pathology and International Vision Correction Research Centre (IVCRC), Department of Ophthalmology, University of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany;David J Apple International Laboratory for Ocular Pathology and International Vision Correction Research Centre (IVCRC), Department of Ophthalmology, University of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany;Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Kangbuk Samsung Hospital, Pyeong-dong, Jongno-gu, Seoul, South Korea;Eye Clinic, Klinikum rechts der Isar der Technischen Universität München, Munich, Germany; | |
关键词: Aspheric IOL; MTF; Optical quality; Strehl ratio; Spherical IOL; | |
DOI : 10.1186/s12886-017-0407-5 | |
received in 2016-11-05, accepted in 2017-01-27, 发布年份 2017 | |
来源: Springer | |
【 摘 要 】
BackgroundTo analyse objective optical properties of the spherical and aspheric design of the same intraocular lens (IOL) model using optical bench analysis.MethodsThis study entailed a comparative analysis of 10 spherical C-flex 570 C and 10 aspheric C-flex 970 C IOLs (Rayner Intraocular Lenses Ltd., Hove, UK) of 26 diopters [D] using an optical bench (OptiSpheric, Trioptics, Germany). In all lenses, we evaluated the modulation transfer function (MTF) at 50 lp/mm and 100 lp/mm and the Strehl Ratio using a 3-mm (photopic) and 4.5-mm (mesopic) aperture.ResultsAt 50 lp/mm, the MTF values were 0.713/0.805 (C-flex 570 C/C-flex 970 C) for a 3-mm aperture and 0.294/0.591 for a 4.5-mm aperture. At 100 lp/mm, the MTF values were 0.524/0.634 for a 3-mm aperture and 0.198/0.344 for a 4.5-mm aperture. The Strehl Ratio was 0.806/0.925 and 0.237/0.479 for a 3-mm and 4.5-mm aperture respectively. A Mann–Whitney U test revealed all intergroup differences to be statistically significant (p < 0.01).ConclusionThe aspheric IOL design achieved higher MTF values than the spherical design of the same IOL for both apertures. Moreover, the differences between the two designs of the IOL were more prominent for larger apertures. This suggests that the evaluated IOL provides enhanced optical quality to patients with larger pupils or working under mesopic conditions.
【 授权许可】
CC BY
© The Author(s). 2017
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202311095094104ZK.pdf | 585KB | download |
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]