BMC Pediatrics
Validity and reliability of electroacoustic probe for diagnosis of developmental dysplasia of the hip
Research Article
Vicente Beltran-Campos1  Silvia del Carmen Delgado-Sandoval2  Nicolas Padilla-Raygoza2  Georgina Olvera-Villanueva2  Teodoro Cordova-Fraga3  Modesto Antonio Sosa-Aquino3 
[1] Department of Clinical Nursing, Division of Health Sciences and Engineering, Campus Celaya Salvatierra, University of Guanajuato, Av. Ing. Javier Barros Sierra 201, 38140, Celaya, Guanajuato, México;Department of Nursing and Obstetrics, Division of Health Sciences and Engineering, Campus Celaya Salvatierra, University of Guanajuato, Mutualismo 303, 38060, Celaya, Guanajuato, México;Department of Physics, Division of Sciences and Engineering, Campus Leon, University of Guanajuato, Lomas del Bosque 103, 37150, Leon, Guanajuato, Mexico;
关键词: Developmental dysplasia of the hip;    Newborns;    Sound transmission;    Ultrasound Graf technique;   
DOI  :  10.1186/s12887-017-0903-z
 received in 2016-01-31, accepted in 2017-06-08,  发布年份 2017
来源: Springer
PDF
【 摘 要 】

BackgroundSound transmission is used in the diagnosis of hip dysplasia since the end of the 80’s. Aim of this study is to quantify the validity and reliability of electroacoustic probe for the diagnosis of hip dysplasia in neonates.MethodsDiagnostic study included neonates aged 4–28 days, whose parents signed an informed consent. The probe was used three times for comparative sound transmission and with extension/flexion; hip ultrasound was performed with Graf technique as gold standard. Kappa was determined for intraobserver and interobserver reliability; validity was calculated with sensitivity, specificity, and predictive values.Results100 neonates were included. For the comparative sound transmission, 0.80 and 0.81 Kappa were obtained for the intraobserver and interobserver respectively; with extension/flexion, Kappa 0.98 and 0.95 were obtained for the intraobserver and interobserver respectively. With comparative sound transmission, 44.8%, 97.7%, 76.5% and 91.3% for sensitivity, specificity, positive and negative predictive values, respectively; with extension/flexion test, the sensitivity, specificity, positive and negative predictive values: 82.8%, 99.4%, 96.0%, and 97.1%, respectively.ConclusionThe electroacoustic probe is moderate valid and reliable for the diagnosis of developmental dysplasia of the hip.Trial registrationOpen Science framework https://osf.io/kpf5s/?view_only=0a9682c6w1c842ad8e1d9a66e8dcf038

【 授权许可】

CC BY   
© The Author(s). 2017

【 预 览 】
附件列表
Files Size Format View
RO202311094494372ZK.pdf 438KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  文献评价指标  
  下载次数:5次 浏览次数:1次