期刊论文详细信息
BMC Ophthalmology
Acute clouding of trifocal lens during implantation: a case report
Case Report
Qian Liu1  Suhua Zhang1  Weifang Cao1  Xiaogang Wang1  Yading Jia1 
[1] Shanxi Eye Hospital, No. 100 Fudong Street, 030002, Taiyuan, Shanxi, People’s Republic of China;
关键词: AT LISA tri 839mp;    Opacification;    Cloudy;   
DOI  :  10.1186/s12886-017-0636-7
 received in 2017-07-12, accepted in 2017-11-28,  发布年份 2017
来源: Springer
PDF
【 摘 要 】

BackgroundIntraoperative IOLs clouding of several kinds of hydrophilic acrylic intraocular lenses (IOLs) have been reported due to temperature changes. This phenomenon reported previously occurred in cold countries and during the winter months. However, no clinical case was reported about trifocal IOL opacification during operation. We report a case in which acute opacification of the optical region occurred simultaneously when AT LISA tri 839mp(Carl Zeiss) was implanted into the eye.Case presentationA 79-year-old woman with a cortex and nucleus cataract was scheduled to undergo right eye phacoemulsification assisted by femtosecond technique. The trifocal lens (AT LISA tri 839mp), which is made of hydrophilic acrylic (25%) with hydrophobic surface properties, was chosen for implantation. As the IOL was implanted into the eye, it became cloudy immediately. Then it was replaced by another AT LISA tri 839mp, which was transferred from lens company outside, the same phenomenon was observed. These two lenses underwent the same temperature fluctuation from cold outside to operating room. Finally, a ZCB00 (Allergan) was implanted.ConclusionsThe acute intraoperative clouding of trifocal lens(AT LISA tri 839mp) results from fluctuation of temperature should be noticed.

【 授权许可】

CC BY   
© The Author(s). 2017

【 预 览 】
附件列表
Files Size Format View
RO202311091604306ZK.pdf 758KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  文献评价指标  
  下载次数:2次 浏览次数:2次