【 摘 要 】

BackgroundRecently, 25-gauge vitrectomy has become more popular. However, most still perform the surgery in pediatric patients without the use of the trocars and cannulas as in adult vitrectomies.MethodsWe described a simple modification using adult 25-gauge cannulas and 270-silicone watzke sleeves, enabling these instruments to be used in pediatric cases. The sleeve is cut into segments of 2 mm in length, and then introduced up the shaft of the 25-gauge trocar. One is introduced first, and a second one is introduced on top of the first one. This secures the two sleeves on the shaft of the trocar, such that they act as a spacer. The effective shaft of the trocar was then reduced to 2 mm in length.ResultsThis method enabled successful surgery in two cases.ConclusionsThis allows the adaptation of the standard 25-gauge system for pediatric cases with only the slightest modification needed.