| BMC Ophthalmology | |
| A simple modification to the 25-gauge trocar and cannula system for retinopathy of prematurity related lens-sparing vitrectomy | |
| Technical Advance | |
| Lawrence P. Iu1  Connie H. Lai1  Ian Y. Wong1  | |
| [1] Department of Ophthalmology, LKS Faculty of Medicine, The University of Hong Kong, Room 301, Level 3, Block B, Cyberport 4, Pokfulam, Hong Kong; | |
| 关键词: Retinal Detachment; Pediatric Case; Luminal Diameter; Vitreous Cavity; Vitrectomy System; | |
| DOI : 10.1186/s12886-016-0214-4 | |
| received in 2015-05-11, accepted in 2016-04-05, 发布年份 2016 | |
| 来源: Springer | |
PDF
|
|
【 摘 要 】
BackgroundRecently, 25-gauge vitrectomy has become more popular. However, most still perform the surgery in pediatric patients without the use of the trocars and cannulas as in adult vitrectomies.MethodsWe described a simple modification using adult 25-gauge cannulas and 270-silicone watzke sleeves, enabling these instruments to be used in pediatric cases. The sleeve is cut into segments of 2 mm in length, and then introduced up the shaft of the 25-gauge trocar. One is introduced first, and a second one is introduced on top of the first one. This secures the two sleeves on the shaft of the trocar, such that they act as a spacer. The effective shaft of the trocar was then reduced to 2 mm in length.ResultsThis method enabled successful surgery in two cases.ConclusionsThis allows the adaptation of the standard 25-gauge system for pediatric cases with only the slightest modification needed.