期刊论文详细信息
Journal of Orthopaedics and Traumatology
Glenoid baseplate fixation using hybrid configurations of locked and unlocked peripheral screws
Original Article
Cooper Wilson1  Nathan T. Formaini2  Jonathan C. Levy2  Nathan G. Everding2  Aniruddh N. Nayak3  Brandon G. Santoni3 
[1]Department of Orthopaedics and Sports Medicine, University of South Florida, Tampa, USA
[2]Holy Cross Orthopedic Institute, 5597 North Dixie Highway, 33334, Fort Lauderdale, FL, USA
[3]Phillip Spiegel Orthopaedic Research Laboratory, Foundation for Orthopaedic Research and Education, Tampa, USA
关键词: Reverse shoulder arthroplasty;    Glenosphere;    Glenoid baseplate;    Hybrid screw fixation;    Micromotion;   
DOI  :  10.1007/s10195-016-0438-3
 received in 2016-03-15, accepted in 2016-11-28,  发布年份 2017
来源: Springer
PDF
【 摘 要 】
BackgroundThe use of peripheral locked screws has reduced glenoid baseplate failure rates in reverse shoulder arthroplasty. However, situations may arise when one or more non-locked screws may be preferred. We aimed to determine if different combinations of locked and non-locked screws significantly alter acute glenoid baseplate fixation in a laboratory setting.Materials and methodsTwenty-eight polyurethane trabecular bone surrogates were instrumented with a center screw-type glenoid baseplate and fixated with various combinations of peripheral locked and non-locked screws (1-, 2-, 3- and 4-locked con). Each construct was tested through a 55° arc of abduction motion generating compressive and shear forces across the glenosphere. Baseplate micromotion (μm) was recorded throughout 10,000 cycles for each model.ResultsAll constructs survived 10,000 cycles of loading without catastrophic failure. One test construct in the 1-locked fixation group exhibited a measured micromotion >150 μm (177.6 μm). At baseline (p > 0.662) and following 10,000 cycles (p > 0.665), no differences were observed in baseplate micromotion for screw combinations that included one, two, three and four peripheral locked screws. The maximum difference in measured micromotion between the extremes of groups (1-locked and 4-locked) was 29 µm.ConclusionsHybrid peripheral screw fixation using combinations of locked and non-locked screws provides secure glenoid baseplate fixation using a polyurethane bone substitute model. Using a glenosphere with a 10-mm lateralized center of rotation, hybrid baseplate fixation maintains micromotion below the necessary threshold for bony ingrowth.Level of EvidenceN/A/, basic science investigation.
【 授权许可】

CC BY   
© The Author(s) 2016

【 预 览 】
附件列表
Files Size Format View
RO202310134885085ZK.pdf 1061KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  文献评价指标  
  下载次数:0次 浏览次数:0次