A proper protocol for routine 18F-FDG uEXPLORER total-body PET/CT scans
Original Research
Jinmei Zhong1  Lijuan Wang1  Kemin Zhou1  Yanjiang Han1  Yanchao Huang1  Huiran Hu1  Li Jiang1  Li Chen1  Hubing Wu1  Hongyan Sun2 
[1] Nanfang PET Center, Nanfang Hospital, Southern Medical University, 1838 Guangzhou Avenue North, 510515, Guangzhou, Guangdong Province, People’s Republic of China;United Imaging Healthcare, Shanghai, People’s Republic of China;
关键词: F-FDG;    uEXPLORER total-body PET/CT;    Image quality;    Scan time;    Injected radioactivity;    Reconstruction algorithm;   
DOI  :  10.1186/s40658-023-00573-4
 received in 2022-08-04, accepted in 2023-08-08,  发布年份 2023
来源: Springer
PDF
【 摘 要 】

BackgroundConventional clinical PET scanners typically have an axial field of view (AFOV) of 15–30 cm, resulting in limited coverage and relatively low photon detection efficiency. Taking advantage of the development of long-axial PET/CT, the uEXPLORER PET/CT scanner with an axial coverage of 194 cm increases the effective count rate by approximately 40 times compared to that of conventional PET scanners. Ordered subset expectation maximization (OSEM) is the most widely used iterative algorithm in PET. The major drawback of OSEM is that the iteration process must be stopped before convergence to avoid image degradation due to excessive noise. A new Bayesian penalized-likelihood iterative PET reconstruction, named HYPER iterative, was developed and is now available on the uEXPLORER total-body PET/CT, which incorporates a noise control component by using a penalty function in each iteration and finds the maximum likelihood solution through repeated iterations. To date, its impact on lesion visibility in patients with a full injected dose or half injected dose is unclear. The goal of this study was to determine a proper protocol for routine 18F-FDG uEXPLORER total-body PET/CT scans.ResultsThe uEXPLORER total-body PET/CT images reconstructed using both OSEM and HYPER iterative algorithms of 20 tumour patients were retrospectively reviewed. The quality of the 5 min PET image was excellent (score 5) for all of the dose and reconstruction methods. Using the HYPER iterative method, the PET images reached excellent quality at 1 min with full-dose PET and at 2 min with half-dose PET. The PET image reached a similar excellent quality at 2 min with a full dose and at 3 min with a half dose using OSEM. The noise in the OSEM reconstruction was higher than that in the HYPER iterative. Compared to OSEM, the HYPER iterative had a slightly higher SUVmax and TBR of the lesions for large positive lesions (≥ 2 cm) (SUVmax: up to 9.03% higher in full dose and up to 12.52% higher in half dose; TBR: up to 8.69% higher in full dose and up to 23.39% higher in half dose). For small positive lesions (≤ 10 mm), the HYPER iterative had an obviously higher SUVmax and TBR of the lesions (SUVmax: up to 45.21% higher in full dose and up to 74.96% higher in half dose; TBR: up to 44.91% higher in full dose and up to 93.73% higher in half dose).ConclusionsA 1 min scan with a full dose and a 2 min scan with a half dose are optimal for clinical diagnosis using the HYPER iterative and 2 min and 3 min for OSEM. For quantification of the small lesions, HYPER iterative reconstruction is preferred.

【 授权许可】

CC BY   
© Springer Nature Switzerland AG 2023

【 预 览 】
附件列表
Files Size Format View
RO202310115307499ZK.pdf 2680KB PDF download
12951_2023_2095_Article_IEq12.gif 1KB Image download
Fig. 5 284KB Image download
12888_2023_5142_Article_IEq13.gif 1KB Image download
13690_2023_1170_Article_IEq76.gif 1KB Image download
13690_2023_1170_Article_IEq205.gif 1KB Image download
Fig. 3 281KB Image download
13063_2023_7612_Figx_HTML.png 4KB Image download
Fig. 2 5714KB Image download
【 图 表 】

Fig. 2

13063_2023_7612_Figx_HTML.png

Fig. 3

13690_2023_1170_Article_IEq205.gif

13690_2023_1170_Article_IEq76.gif

12888_2023_5142_Article_IEq13.gif

Fig. 5

12951_2023_2095_Article_IEq12.gif

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  • [38]
  • [39]
  • [40]
  • [41]
  • [42]
  • [43]
  • [44]
  • [45]
  • [46]
  • [47]
  文献评价指标  
  下载次数:8次 浏览次数:0次