Capture threshold of bipolar and unipolar pacing of left ventricle via coronary sinus branch: longitudinal study
Cardiovascular Medicine
Jakub Simka1  Ludek Haman1  Petr Parizek1  Josef Hanus2  Jan Kremlacek2  Jakub Stritecky3  Hana Stritecka4 
[1] 1st Department of Internal Medicine – Cardioangiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic;Department of Medical Biophysics, Faculty of Medicine in Hradec Kralove, Charles University, Hradec Králové, Czech Republic;Department of Medical Biophysics, Faculty of Medicine in Hradec Kralove, Charles University, Hradec Králové, Czech Republic;1st Department of Internal Medicine – Cardioangiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic;Department of Military Internal Medicine and Military Hygiene, Faculty of Health Sciences, University of Defence, Hradec Králové, Czech Republic;
关键词: quadripolar lead;    pacing capture threshold energy;    pacing polarity;    steroid eluting lead;    resynchronisation therapy;    battery longevity;   
DOI  :  10.3389/fcvm.2023.1096538
 received in 2022-11-12, accepted in 2023-04-18,  发布年份 2023
来源: Frontiers
PDF
【 摘 要 】

IntroductionThe aim of this paper is to first monitor the changes in the capture threshold of endovascularly placed leads for left ventricle pacing, second to compare the pacing configurations, and third to verify the effect of Steroid elution for endovascular leads.Sample and MethodThe study included 202 consecutive single centre patients implanted with the Quartet™ lead (St. Jude Medical). The capture threshold and related lead parameters were tested during implantation, on the day of the patient’s discharge, and 3, 9, and 15 months after implantation. The electrical energy corresponding to the threshold values for inducing ventricular contraction was recorded for subgroups of patients with bipolar and pseudo-unipolar pacing vectors and electrodes equipped with and without a slow-eluting steroids. The best setting for the resynchronization effect was generally chosen. Capture threshold was taken as a selection criterion only if there were multiple options with (expected) similar resynchronization effect.Results and DiscussionThe measurements showed that the ratio of threshold energies of UNI vs. BI was 5× higher (p < 0.001) at implantation. At the end of the follow-up, it dropped to 2.6 (p = 0.012). The steroid effect in BI vectors was caused by a double capture threshold in the NSE group compared to the SE group (p < 0.001), increased by approximately 2.5 times (p < 0.001). The study concludes that after a larger initial increase in the capture threshold, the leads showed a gradual increase in the entire set. As a result, the bipolar threshold energies increase, and the pseudo-unipolar energies decrease. Since bipolar vectors require a significantly lower pacing energy, battery life of the implanted device would improve. When evaluating the steroid elution of bipolar vectors, we observe a significant positive effect of a gradual increase of the threshold energy.

【 授权许可】

Unknown   
© 2023 Stritecky, Kremlacek, Hanus, Haman, Stritecka, Simka and Parizek.

【 预 览 】
附件列表
Files Size Format View
RO202310101070534ZK.pdf 1391KB PDF download
  文献评价指标  
  下载次数:1次 浏览次数:0次