Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer
Research
Alberto Domínguez-Vicent1  Amanda Dalin1  Rune Brautaset1  Abinaya Priya Venkataraman1  Robert Montés-Micó2 
[1] Division of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, 171 77, Stockholm, Sweden;Optics and Optometry and Vision Sciences Department, University of Valencia, Valencia, Spain;
关键词: Optical biometry;    Swept-source optical coherence tomography;    Low-coherence interferometry;    Repeatability;    Agreement;   
DOI  :  10.1186/s40662-023-00343-4
 received in 2022-12-12, accepted in 2023-04-21,  发布年份 2023
来源: Springer
PDF
【 摘 要 】

BackgroundTo evaluate the repeatability of a fully automated swept-source optical coherence tomography (SS-OCT) and its agreement with an optical low coherence reflectometry (OLCR) for several biometric parameters.MethodsIn this study, 74 eyes of 74 patients were measured using the Eyestar 900 SS-OCT and Lenstar LS 900 OLCR. Flat keratometry (K1) and steep keratometry (K2), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and axial length (AL) were measured three times with each device. The repeatability was analyzed with the intrasubject standard deviation, coefficient of variability (CoV), and coefficient of repeatability (CoR) for each instrument. The agreement between the instruments was evaluated with Bland-Altman analysis.ResultsK1, K2 and CCT CoV values were < 0.2%, < 0.4% and < 0.55%, respectively. Higher CoV values were found for ACD and LT ranging from 0.56% to 1.74%. The lowest CoV values were found for the AL measurements (0.03% and 0.06% for the Eyestar 900 and the Lenstar LS 900, respectively). AL measurements provided the highest repeatability, measured with both CoV and CoR values, and the CCT was the parameter with the lowest repeatability. The CCT and LT measurements were statistically significant between the two biometers (P < 0.001). The interval of the limits of agreement was < 0.6 D for K1 and K2, 15.78 µm for CCT, 0.21 mm for ACD, 0.34 mm for LT, and 0.08 mm for AL.ConclusionsBoth biometers provide repeatable measurements for the different parameters analyzed and can be used interchangeably.

【 授权许可】

CC BY   
© The Author(s) 2023

【 预 览 】
附件列表
Files Size Format View
RO202309078698152ZK.pdf 1490KB PDF download
40517_2023_252_Article_IEq123.gif 1KB Image download
Fig. 1 1566KB Image download
Fig. 2 415KB Image download
40517_2023_252_Article_IEq134.gif 1KB Image download
12302_2023_754_Article_IEq4.gif 1KB Image download
Fig. 2 518KB Image download
Fig. 1 16KB Image download
MediaObjects/12951_2023_1959_MOESM3_ESM.tif 995KB Other download
【 图 表 】

Fig. 1

Fig. 2

12302_2023_754_Article_IEq4.gif

40517_2023_252_Article_IEq134.gif

Fig. 2

Fig. 1

40517_2023_252_Article_IEq123.gif

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  文献评价指标  
  下载次数:7次 浏览次数:5次