BMC Ophthalmology
Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes
Research
Yongdong Lin1  Zifeng Zhao1  Hongxi Wang1  Guihua Zhang1  Yongqun Xiong1 
[1] Joint Shantou International Eye Center of Shantou University, The Chinese University of Hong Kong, Dong xia Road, Shantou, Guangdong Province, People’s Republic of China;
关键词: Ocular biometry;    Silicone oil-filled eyes;    Retinal disease;    Swept-source optical coherence tomography;   
DOI  :  10.1186/s12886-023-02962-w
 received in 2022-09-15, accepted in 2023-05-05,  发布年份 2023
来源: Springer
PDF
【 摘 要 】

PurposeTo evaluate a new non-contact instrument (OA-2000) measuring the ocular biometry parameters of silicone oil (SO)-filled aphakic eyes, as compared with IOLMaster 700.MethodsForty SO-filled aphakic eyes of 40 patients were enrolled in this cross-sectional clinical trial. The axial length (AL), central corneal thickness (CCT), keratometry ((flattest keratometry) Kf and (steep keratometry, 90° apart from Kf) Ks), and axis of the Kf (Ax1) were measured with OA-2000 and IOLMaster 700. The coefficient of variation (CoV) was calculated to assess the repeatability. The correlation was evaluated by the Pearson coefficient. Bland-Altman analysis and paired t test were used to analyze the agreements and differences of parameters measured by the two devices, respectively.ResultsThe mean AL obtained with the OA-2000 was 23.57 ± 0.93 mm (range: 21.50 to 25.68 mm), and that obtained with the IOLMaster 700 was 23.69 ± 0.94 mm (range: 21.85 to 25.86 mm), resulting in a mean offset of 0.124 ± 0.125 mm (p < 0.001). The mean offset of CCT measured by OA-2000 and IOLMaster 700 was 14.6 ± 7.5 μm (p < 0.001). However, the Kf, Ks and Ax1 values from the two devices were comparable (p > 0.05). All the measured parameters of the two devices showed strong linear correlations (all r ≥ 0.966). The Bland-Altman analysis showed a narrow 95% limits of agreement (LoA) of Kf, Ks and AL, but 95%LoA of CCT and Ax1 was wide, which were − 29.3 ~ 0.1 μm and-25.9 ~ 30.7°respectively. The CoVs of the biometric parameters obtained with OA-2000 were lower than 1%.ConclusionIn SO-filled aphakic eyes, the ocular parameters (including AL, Kf, Ks, Ax1, and CCT) measured by the OA-2000 and IOLMaster 700 had a good correlation. Two devices had an excellent agreement on ocular biometric measurements of Kf, Ks and AL. The OA-2000 provided excellent repeatability of ocular parameters in SO-filled aphakic eyes.

【 授权许可】

CC BY   
© The Author(s) 2023

【 预 览 】
附件列表
Files Size Format View
RO202308156686588ZK.pdf 1668KB PDF download
MediaObjects/12888_2023_4867_MOESM8_ESM.docx 17KB Other download
41116_2023_36_Article_IEq413.gif 1KB Image download
Fig. 1 835KB Image download
41116_2023_36_Article_IEq417.gif 1KB Image download
41116_2023_36_Article_IEq419.gif 1KB Image download
【 图 表 】

41116_2023_36_Article_IEq419.gif

41116_2023_36_Article_IEq417.gif

Fig. 1

41116_2023_36_Article_IEq413.gif

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  • [38]
  • [39]
  文献评价指标  
  下载次数:2次 浏览次数:0次