European Radiology Experimental
Improved visualization of the bone-implant interface and osseointegration in ex vivo acetabular cup implants using photon-counting detector CT
Original Article
Erik Tesselaar1  Jörg Schilcher2  Ronald Booij3  Edwin H. G. Oei4  Mischa Woisetschläger5 
[1] Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;Department of Medical Radiation Physics, and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden;Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;Department of Orthopedics, Department of Clinical and Experimental Medicine, Linköping University, Linköping, Sweden;Wallenberg Center for Molecular Medicine, Linköping University, Linköping, Sweden;Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;Department of Radiology & Nuclear Medicine, Erasmus Medical Center, Rotterdam, The Netherlands;Department of Radiology & Nuclear Medicine, Erasmus Medical Center, Rotterdam, The Netherlands;Department of Radiology and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden;Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;
关键词: Arthroplasty;    Joint prosthesis;    Osseointegration;    Osteoarthritis (hip);    Tomography (x-ray computed);   
DOI  :  10.1186/s41747-023-00335-y
 received in 2023-01-06, accepted in 2023-02-28,  发布年份 2023
来源: Springer
PDF
【 摘 要 】

BackgroundSuccessful osseointegration of joint replacement implants is required for long-term implant survival. Accurate assessment of osseointegration could enable clinical discrimination of failed implants from other sources of pain avoiding unnecessary surgeries. Photon-counting detector computed tomography (PCD-CT) provides improvements in image resolution compared to conventional energy-integrating detector CT (EID-CT), possibly allowing better visualization of bone-implant-interfaces and osseointegration. The aim of this study was to assess the quality of visualization of bone-implant-interfaces and osseointegration in acetabular cup implants, using PCD-CT compared with EID-CT.MethodsTwo acetabular implants (one cemented, one uncemented) retrieved during revision surgery were scanned using PCD-CT and EID-CT at equal radiation dose. Images were reconstructed using different reconstruction kernels and iterative strengths. Delineation of the bone-implant and bone-cement-interface as an indicator of osseointegration was scored subjectively for image quality by four radiologists on a Likert scale and assessed quantitatively.ResultsDelineation of bone-implant and bone-cement-interfaces was better with PCD-CT compared with EID-CT (p ≤ 0.030). The highest ratings were given for PCD-CT at sharper kernels for the cemented cup (PCD-CT, median 5, interquartile range 4.25–5.00 versus EID-CT, 3, 2.00–3.75, p < 0.001) and the uncemented cup (5, 4.00–5.00 versus 2, 2–2, respectively, p < 0.001). The bone-implant-interface was 35–42% sharper and the bone-cement-interface was 28–43% sharper with PCD-CT compared with EID-CT, depending on the reconstruction kernel.ConclusionsPCD-CT might enable a more accurate assessment of osseointegration of orthopedic joint replacement implants.Key points• The bone-implant interface ex vivo showed superior visualization using photon-counting detector computed tomography (PCD-CT) compared to energy-integrating detector computed tomography.• Harder reconstruction kernels in PCD-CT provide sharper images with lower noise levels.• These improvements in imaging might make it possible to visualize osseointegration in vivo.

【 授权许可】

CC BY   
© The Author(s) 2023

【 预 览 】
附件列表
Files Size Format View
RO202308153534563ZK.pdf 2107KB PDF download
41116_2023_36_Article_IEq682.gif 1KB Image download
MediaObjects/12888_2023_4761_MOESM1_ESM.docx 55KB Other download
41116_2023_36_Article_IEq786.gif 1KB Image download
MediaObjects/40249_2023_1063_MOESM8_ESM.docx 62KB Other download
40517_2023_256_Article_IEq33.gif 1KB Image download
40517_2023_256_Article_IEq34.gif 1KB Image download
40517_2023_256_Article_IEq35.gif 1KB Image download
【 图 表 】

40517_2023_256_Article_IEq35.gif

40517_2023_256_Article_IEq34.gif

40517_2023_256_Article_IEq33.gif

41116_2023_36_Article_IEq786.gif

41116_2023_36_Article_IEq682.gif

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  文献评价指标  
  下载次数:12次 浏览次数:2次