| European Radiology Experimental | |
| Improved visualization of the bone-implant interface and osseointegration in ex vivo acetabular cup implants using photon-counting detector CT | |
| Original Article | |
| Erik Tesselaar1  Jörg Schilcher2  Ronald Booij3  Edwin H. G. Oei4  Mischa Woisetschläger5  | |
| [1] Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;Department of Medical Radiation Physics, and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden;Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;Department of Orthopedics, Department of Clinical and Experimental Medicine, Linköping University, Linköping, Sweden;Wallenberg Center for Molecular Medicine, Linköping University, Linköping, Sweden;Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden;Department of Radiology & Nuclear Medicine, Erasmus Medical Center, Rotterdam, The Netherlands;Department of Radiology & Nuclear Medicine, Erasmus Medical Center, Rotterdam, The Netherlands;Department of Radiology and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden;Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden; | |
| 关键词: Arthroplasty; Joint prosthesis; Osseointegration; Osteoarthritis (hip); Tomography (x-ray computed); | |
| DOI : 10.1186/s41747-023-00335-y | |
| received in 2023-01-06, accepted in 2023-02-28, 发布年份 2023 | |
| 来源: Springer | |
PDF
|
|
【 摘 要 】
BackgroundSuccessful osseointegration of joint replacement implants is required for long-term implant survival. Accurate assessment of osseointegration could enable clinical discrimination of failed implants from other sources of pain avoiding unnecessary surgeries. Photon-counting detector computed tomography (PCD-CT) provides improvements in image resolution compared to conventional energy-integrating detector CT (EID-CT), possibly allowing better visualization of bone-implant-interfaces and osseointegration. The aim of this study was to assess the quality of visualization of bone-implant-interfaces and osseointegration in acetabular cup implants, using PCD-CT compared with EID-CT.MethodsTwo acetabular implants (one cemented, one uncemented) retrieved during revision surgery were scanned using PCD-CT and EID-CT at equal radiation dose. Images were reconstructed using different reconstruction kernels and iterative strengths. Delineation of the bone-implant and bone-cement-interface as an indicator of osseointegration was scored subjectively for image quality by four radiologists on a Likert scale and assessed quantitatively.ResultsDelineation of bone-implant and bone-cement-interfaces was better with PCD-CT compared with EID-CT (p ≤ 0.030). The highest ratings were given for PCD-CT at sharper kernels for the cemented cup (PCD-CT, median 5, interquartile range 4.25–5.00 versus EID-CT, 3, 2.00–3.75, p < 0.001) and the uncemented cup (5, 4.00–5.00 versus 2, 2–2, respectively, p < 0.001). The bone-implant-interface was 35–42% sharper and the bone-cement-interface was 28–43% sharper with PCD-CT compared with EID-CT, depending on the reconstruction kernel.ConclusionsPCD-CT might enable a more accurate assessment of osseointegration of orthopedic joint replacement implants.Key points• The bone-implant interface ex vivo showed superior visualization using photon-counting detector computed tomography (PCD-CT) compared to energy-integrating detector computed tomography.• Harder reconstruction kernels in PCD-CT provide sharper images with lower noise levels.• These improvements in imaging might make it possible to visualize osseointegration in vivo.
【 授权许可】
CC BY
© The Author(s) 2023
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202308153534563ZK.pdf | 2107KB | ||
| 41116_2023_36_Article_IEq682.gif | 1KB | Image | |
| MediaObjects/12888_2023_4761_MOESM1_ESM.docx | 55KB | Other | |
| 41116_2023_36_Article_IEq786.gif | 1KB | Image | |
| MediaObjects/40249_2023_1063_MOESM8_ESM.docx | 62KB | Other | |
| 40517_2023_256_Article_IEq33.gif | 1KB | Image | |
| 40517_2023_256_Article_IEq34.gif | 1KB | Image | |
| 40517_2023_256_Article_IEq35.gif | 1KB | Image |
【 图 表 】
40517_2023_256_Article_IEq35.gif
40517_2023_256_Article_IEq34.gif
40517_2023_256_Article_IEq33.gif
41116_2023_36_Article_IEq786.gif
41116_2023_36_Article_IEq682.gif
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]
- [28]
- [29]
- [30]
- [31]
- [32]
PDF