| Biomechanical evaluation of locked plating fixation for unstable femoral neck fractures | |
| article | |
| E. Bliven1  S. Sandriesser1  P. Augat1  C. von Rüden3  S. Hackl1  | |
| [1] Institute for Biomechanics;Paracelsus Medical University;Department for Trauma Surgery;Institute for Biomechanics, Paracelsus Medical University | |
| 关键词: Biomechanics; Osteosynthesis; Femoral neck; Fracture fixation; Plate; | |
| DOI : 10.1302/2046-3758.96.BJR-2019-0331.R1 | |
| 学科分类:骨科学 | |
| 来源: British Editorial Society Of Bone And Joint Surgery | |
PDF
|
|
【 摘 要 】
AimsEvaluate if treating an unstable femoral neck fracture with a locking plate and spring-loaded telescoping screw system would improve construct stability compared to gold standard treatment methods.MethodsA 31B2 Pauwels’ type III osteotomy with additional posterior wedge was cut into 30 fresh-frozen femur cadavers implanted with either: three cannulated screws in an inverted triangle configuration (CS), a sliding hip screw and anti-rotation screw (SHS), or a locking plate system with spring-loaded telescoping screws (LP). Dynamic cyclic compressive testing representative of walking with increasing weight-bearing was applied until failure was observed. Loss of fracture reduction was recorded using a high-resolution optical motion tracking system.ResultsLP constructs demonstrated the highest mean values for initial stiffness and failure load. LP and SHS constructs survived on mean over 50% more cycles and to loads 450 N higher than CS. During the early stages of cyclic loading, mean varus collapse of the femoral head was 0.5° (SD 0.8°) for LP, 0.7° (SD 0.7°) for SHS, and 1.9° (SD 2.3°) for CS (p = 0.071). At 30,000 cycles (1,050 N) mean femoral neck shortening was 1.8 mm (SD 1.9) for LP, 2.0 mm (SD 0.9) for SHS, and 3.2 mm (SD 2.5) for CS (p = 0.262). Mean leg shortening at construct failure was 4.9 mm (SD 2.7) for LP, 8.9 mm (SD 3.2) for SHS, and 7.0 mm (SD 4.3) for CS (p = 0.046).ConclusionUse of the LP system provided similar (hip screw) or better (cannulated screws) biomechanical performance as the current gold standard methods suggesting that the LP system could be a promising alternative for the treatment of unstable fractures of the femoral neck.
【 授权许可】
CC BY-NC
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202307110000578ZK.pdf | 1231KB |
PDF