期刊论文详细信息
Saudi Endodontic Journal
Influence of different cross-section on cyclic fatigue resistance of two nickel–titanium rotary instruments with same heat treatment: An in vitro study
article
Dario Di Nardo1  Gianluca Gambarini1  Marco Seracchiani1  Alessandro Mazzoni1  Alessio Zanza1  Andrea Del Giudice1  Maurilio D'Angelo1  LucaTestarelli1 
[1] Department of Oral and Maxillo-Facial Sciences, “Sapienza” University of Rome
关键词: Cross-section;    cyclic fatigue;    endodontics;    heat treatment;    nickel–titanium rotary instruments;   
DOI  :  10.4103/sej.sej_124_19
学科分类:口腔科学
来源: Medknow
PDF
【 摘 要 】

Introduction: The aim of this study was to investigate the role of the cross-sectional design surface in improving the cyclic fatigue resistance by comparing two nickel–titanium (NiTi) rotary instruments with the same heat-treatment and grinding procedure. Materials and Methods: A total of 40 new NiTi instruments (20 S-One 25.06 taper and 20 AF Blue S4 25.06 taper) were tested in the present study. Both instruments were rotated at 300 rpm and with a torque setting of 2 Ncm) using an endodontic motor in the same artificial canal (90° angle of curvature and 3 mm radius). A Student's t-test was performed to determinate the differences in terms of time to fracture (TtF) and the length of the fractured portion between the two different instruments. Significance was set at P < 0.05. Results: The differences in terms of TtF for cyclic fatigue resistance between the two files were statistically significant (P < 0.05). In terms of fractured length, not statistically significant differences were found (P 0.05). Scanning electron microscopy images showed patterns of cyclic fatigue striations. Conclusion: The results of this study confirm and enhance the role of cross-section in increasing the cyclic fatigue resistance of NiTi rotary instruments.

【 授权许可】

CC BY-NC-SA   

【 预 览 】
附件列表
Files Size Format View
RO202307100000807ZK.pdf 2087KB PDF download
  文献评价指标  