Article
关键词: COMPRESSION-DECOMPRESSION CPR;    CARDIOPULMONARY-RESUSCITATION;    THRESHOLD VALVE;    PROGRESS REPORT;   
DOI  :  10.1161/hc0102.101391
来源: SCIE
【 摘 要 】

Background-This study evaluated the potential for an inspiratory impedance threshold valve (ITV) to improve 24-hour survival and neurological function in a pig model of cardiac arrest. Methods and Results-Using a randomized, prospective, and blinded design, we compared the effects of a sham versus active ITV on 24-hour survival and neurological function. After 6 minutes of ventricular fibrillation (VF), followed by 6 minutes of cardiopulmonary resuscitation (CPR) with either a sham or an active valve, anesthetized pigs received 3 sequential 200-J shocks. If VF persisted, they received epinephrine (0.045 mg/kg), 90 seconds of CPR, and 3 more 200-J shocks. A total of 11 of 20 pigs (55%) in the sham versus 17 of 20 (85%) in the active valve group survived for 24 hours (P<0.05). Neurological scores were significantly higher with the active valve; the cerebral performance score (1 = normal, 5 = brain death) was 2.2+/-0.2 with the sham ITV versus 1.4+/-0.2 with the active valve (P<0.05). A total of 1 of 11 in the sham versus 12 of 17 in the active valve group had completely normal neurological function (P<0.05). Peak end-tidal CO2 (PETCO2) values were significantly higher with the active valve (20.4+/-1.0) than the sham (16.8+/-1.5) (P<0.05). PETCO2 > 18 mm Hg correlated with increased survival (P<0.05). Conclusions-Use of a functional ITV during standard CPR significantly improved 24-hour survival rates and neurological recovery. PETCO2 and systolic blood pressure were also significantly higher in the active valve group. These data support further evaluation of ITV during standard CPR.

【 授权许可】

Free   

  文献评价指标  
  下载次数:0次 浏览次数:6次