Comparison between the CASIA SS-1000 and Pentacam in measuring corneal curvatures and corneal thickness maps
Research
Atsushi Shiraishi1  Alice Z. Chuang2  Robert M. Feldman2  Gene Kim3  Makoto Tsukamoto4  Keiichiro Okamoto4 
[1] Department of Ophthalmology, Ehime University Graduate School of Medicine, Shitsukawa, Toon, Ehime, Japan;Ruiz Department of Ophthalmology and Visual Science, McGovern Medical School at The University of Texas Health Science Center at Houston (UTHealth), 6400 Fannin St., Suite #1800, 77030, Houston, TX, USA;Ruiz Department of Ophthalmology and Visual Science, McGovern Medical School at The University of Texas Health Science Center at Houston (UTHealth), 6400 Fannin St., Suite #1800, 77030, Houston, TX, USA;Department of Ophthalmology, Dell Medical School, Austin, TX, USA;Tomey Corp, Nagoya, Japan;
关键词: CASIA SS-1000;    Pentacam;    Corneal curvature;    Laser refractive surgery;   
DOI  :  10.1186/s12886-023-02768-w
 received in 2022-05-31, accepted in 2023-01-02,  发布年份 2023
来源: Springer
PDF
【 摘 要 】

PurposeTo compare the intra-device repeatability and inter-device reproducibility between two anterior segment imaging instruments, the CASIA SS-1000 (Tomey Corp., Nagoya, Japan) and Pentacam (OCULUS, Arlington, WA) in measuring anterior segment parameters.MethodsSingle-center, prospective clinical trial. Participants ≥20 years of age were included. One eye was randomly selected, each imaged by three CASIA SS-1000 devices and three Pentacam devices by three different examiners. Each photographer operated a pair of devices, one CASIA SS-1000 and one Pentacam. The image order for each participant was determined by a random permutation table. Three images were taken from each device. A total of 18 images were taken for each eye. Ratios of the standard deviations, referenced as (CASIA/Pentacam), were calculated to compare the repeatability and reproducibility of the two imaging instruments.ResultsIn all, 66 participants with a mean age of 46.4 years (±21.7) were enrolled in the study. All repeatability ratios and intra-device variability were less than 1 (anterior corneal curvature: flat = 0.86, steep = 0.85; posterior corneal curvature: flat = 0.43, steep = 0.61; and map: thinnest = 0.22; central = 0.24, 2 mm = 0.26, 4 mm = 0.27, and 6 mm = 0.30). All reproducibility ratios, which measure the inter-device variability, were less than 1 (anterior corneal curvature: flat = 0.58, steep = 0.73; posterior corneal curvature: flat = 0.25, steep = 0.31; and pachymetry map: thinnest = 0.20; central = 0.20; 2 mm = 0.20; 4 mm = 0.19; and 6 mm = 0.22). A ratio of less than 1 indicates that the CASIA SS-1000 has more consistent measurements.ConclusionsThe CASIA SS-1000 was found to have better repeatability and reproducibility compared to the Pentacam for both corneal curvature and pachymetry maps. This greater consistency may require further study to determine whether the decreased variability can be translated into improved clinical results.

【 授权许可】

CC BY   
© The Author(s) 2023

【 预 览 】
附件列表
Files Size Format View
RO202305116704636ZK.pdf 2079KB PDF download
Fig. 3 55KB Image download
MediaObjects/41408_2023_789_MOESM1_ESM.docx 641KB Other download
【 图 表 】

Fig. 3

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  文献评价指标  
  下载次数:11次 浏览次数:0次