Validation of the vault prediction model based on the sulcus-to-sulcus diameter and lens thickness: a 925-eye prospective study
Research
Lie Ma1  Man-Hui Zhu1  E. Song1  You Yuan1  Qiu-Jian Zhu2  Xiao-Ying Xing3 
[1] Department of Ophthalmology, Lixiang Eye Hospital of Soochow University, Suzhou, China;Department of Ophthalmology, The Second Affiliated Hospital of Soochow University, Suzhou, China;Department of Ophthalmology, Lixiang Eye Hospital of Soochow University, Suzhou, China;Suzhou Eye Hospital, Suzhou, China;
关键词: ICL implantation;    Prediction formula;    Vault;    Sulcus to sulcus;    UBM;   
DOI  :  10.1186/s12886-022-02698-z
 received in 2022-04-24, accepted in 2022-11-21,  发布年份 2022
来源: Springer
PDF
【 摘 要 】

BackgroundTo verify the accuracy and stability of the prediction formula based on the ciliary sulcus diameter and lens thickness and to analyse factors influencing the prediction results.MethodsIn total, 925 eyes from 506 subjects were enrolled in this prospective study between July 1, 2020, and June 30, 2021. Subjects were divided into four seasons, each spanning three months. The target vault was set to be between 300 μm and 700 μm according the prediction formula. The actual vault was measured one month postoperatively. The Bland–Altman test, 95% confidence intervals (95% CI) and 95% limits of agreement (95% LoA) were used to evaluate the agreement between the predicted vault and the actual vault. Eyes with absolute prediction errors greater than 300 μm were further analysed.ResultsThe mean predicted vaults for the four seasons were 503 ± 99, 494 ± 96, 481 ± 92 and 502 ± 93 μm, while the mean actual vaults were 531 ± 189, 491 ± 179, 464 ± 179 and 529 ± 162 μm, respectively. The predicted and actual vaults of the overall subjects were 493 ± 95 and 500 ± 180 μm, respectively. Of the 925 eyes, 861 eyes (93.08%), 42 eyes (4.54%), and 22 eyes (2.38%) showed a normal vault, high vault, and low vault, respectively. Bland–Altman plots showed that the mean difference between the actual vault and predicted vault overall (± 95% LoA) was 6.43 ± 176.2 μm (-339 to 352 μm). Three UBM features may lead to large prediction errors (more than 300 μm): wide iris-ciliary angle (ICA), iris concavity and anteriorly positioned ciliary body.ConclusionsThis study demonstrated the accuracy and stability of the prediction formula through the validation of a large sample size and a long time span. Wide ICA, iris concavity and anteriorly positioned ciliary body may have an effect on vault.

【 授权许可】

CC BY   
© The Author(s) 2022

【 预 览 】
附件列表
Files Size Format View
RO202305064811137ZK.pdf 1061KB PDF download
12902_2022_1244_Article_IEq30.gif 1KB Image download
MediaObjects/12974_2022_2668_MOESM6_ESM.tif 1339KB Other download
Fig. 2 331KB Image download
【 图 表 】

Fig. 2

12902_2022_1244_Article_IEq30.gif

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  • [38]
  文献评价指标  
  下载次数:13次 浏览次数:3次