Quantitative Imaging in Medicine and Surgery | |
A miniature U-net for k -space-based parallel magnetic resonance imaging reconstruction with a mixed loss function | |
article | |
Qian Zheng1  Jianying Yuan2  Jiajia Liu2  Lin Xu3  Jingwen Xu4  | |
[1] College of Software Engineering , Zhengzhou University of Light Industry;Control Engineering College , Chengdu University of Information Technology;College of Medical Information Engineering , Chengdu University of Traditional Chinese Medicine;Department of Geriatrics , the General Hospital of Western Theater Command | |
关键词: Magnetic resonance imaging (MRI); U-net; deep learning; parallel imaging; | |
DOI : 10.21037/qims-21-1212 | |
学科分类:外科医学 | |
来源: AME Publications | |
【 摘 要 】
Background: Deep learning-based magnetic resonance imaging (MRI) methods require in most cases a separate dataset with thousands of images for each anatomical site to train the network model. This paper proposes a miniature U-net method for k-space-based parallel MRI where the network model is trained individually for each scan using scan-specific autocalibrating signal data. Methods: The original U-net was tailored with fewer layers and channels, and the network was trained using the autocalibrating signal data with a mixing loss function involving magnitude loss and phase loss. The performance of the proposed method was measured using both phantom and in vivo datasets compared to scan-specific robust artificial-neural-networks for k-space interpolation (RAKI) and generalized autocalibrating partially parallel acquisitions (GRAPPA). Results: The proposed method alleviates aliasing artifacts and reduces noise with an acceleration factor of four for phantom and in vivo data. Compared with RAKI and GRAPPA, the proposed method represents an improvement with a structural similarity index measure of between 0.02 and 0.05 and a peak signal-to-noise ratio (PSNR) of between 0.1 and 3. Conclusions: The proposed method introduces a miniature U-net to reconstruct the missing k-space data, which can provide an optimal trade-off between network performance and requirement of training samples. Experimental results indicate that the proposed method can improve image quality compared with the deep learning-based k-space parallel magnitude resonance imaging method.
【 授权许可】
All Rights reserved
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202303290000480ZK.pdf | 5213KB | download |