期刊论文详细信息
Laryngoscope Investigative Otolaryngology
Effect of inner ear malformations on intraoperative ECAP thresholds and postoperative auditory performance
article
Jeong-Seo Kim AuD, PhD1  Sung Hwa Hong MD, PhD1  Il Joon Moon MD, PhD1 
[1] Hearing Research Laboratory, Samsung Medical Center;Department of Otolaryngology – Head and Neck Surgery, Samsung Changwon Hospital, Sungkyunkwan University School of Medicine;Department of Otolaryngology – Head and Neck Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine
关键词: categories of auditory performance;    cochlear implant;    electrically evoked compound action potential;    inner ear malformation;    neural response telemetry;   
DOI  :  10.1002/lio2.836
学科分类:环境科学(综合)
来源: Wiley
PDF
【 摘 要 】

Objectives This study sought to characterize the influence of inner ear malformations (IEMs) on intraoperative electrically evoked compound action potential (ECAP) and auditory performance to better understand the underlying pathophysiology related to variabilities in cochlear implant (CI) outcomes that individuals with malformed cochlea may present. Methods The medical records of 222 ears implanted with Cochlear Nucleus CI were reviewed. Of the total, 64 ears had radiologic evidence of IEMs, and 158 ears were normal. Individuals with IEMs were grouped based on the severity of anomalies; 38 had mild IEMs (e.g., enlarged vestibular aqueduct, incomplete partition type II, etc.) and 26 had severe IEMs (e.g., cochlear nerve hypoplasia, common cavity, etc.). Intraoperative ECAP thresholds obtained via neural response telemetry (NRT) and the categories of auditory performance (CAP) scores measured at 12 months postoperative were compared and correlated. Results Absent ECAP responses were more apparent in the IEM group. ECAP thresholds were significantly elevated in the severe IEM group, while the mild IEM group had ECAP thresholds comparable to the normal group. The mild IEM group achieved CAP scores similar to the normal control. Patients in the severe IEM group showed significantly lower CAP scores at 12 months postoperative. Significant negative relationships existed between ECAP thresholds and CAP scores obtained from all subjects. Conclusion Measurable ECAP responses and NRT thresholds varied across groups. The inverse relationship between NRT thresholds and CAP scores may suggest that electrophysiological responses measured during surgery may potentially be indicative of postoperative performance in our CI population. Level of Evidence 2b.

【 授权许可】

CC BY|CC BY-NC-ND   

【 预 览 】
附件列表
Files Size Format View
RO202302050004341ZK.pdf 895KB PDF download
  文献评价指标  