Feasibility of an Automatic Ultrasonographic Image Acquisition System Associated With an Artificial Intelligence Algorithm for Real-Time Monitoring of Cardiac Motion During Cardiac Radio-Ablation | |
article | |
Matteo Casula1  Veronica Dusi1  Saskia Camps4  Jérémie Gringet4  Tristan Benoit4  Adriano Garonna4  Roberto Rordorf1  | |
[1] Arrhythmia and Electrophysiology Unit, Fondazione IRCCS Policlinico San Matteo;Section of Cardiology, Department of Molecular Medicine, University of Pavia;Unit of Cardiology;EBAMed SA | |
关键词: cardiac radioablation; motion monitoring; ventricular arrhythmia; echocardiography; artificial intelligence; | |
DOI : 10.3389/fcvm.2022.849234 | |
学科分类:地球科学(综合) | |
来源: Frontiers | |
【 摘 要 】
Background The management of the cardio-respiratory motion of the target and the reduction of the uncertainties related to patient's positioning are two of the main challenges that stereotactic arrhythmia radio-ablation (STAR) has to overcome. A prototype of a system was developed that can automatically acquire and interpret echocardiographic images using an artificial intelligence (AI) algorithm to calculate cardiac displacement in real-time. Methods We conducted a single center study enrolling consecutive patients with a history of ventricular arrhythmias (VA) in order to evaluate the feasibility of this automatic acquisition system. Echocardiographic images were automatically acquired from the parasternal and apical views with a dedicated probe. The system was designed to hold the probe fixed to the chest in the supine position during both free-breathing and short expiratory breath-hold sequences, to simulate STAR treatment. The primary endpoint was the percentage of patients reaching a score ≥2 in a multi-parametric assessment evaluating the quality of automatically acquired images. Moreover, we investigated the potential impact of clinical and demographic characteristics on achieving the primary endpoint. Results We enrolled 24 patients (63 ± 14 years, 21% females). All of them had a history of VA and 21 (88%) had an ICD. Eight patients (33%) had coronary artery disease, 12 (50%) had non-ischemic cardiomyopathy, and 3 had idiopathic VA. Parasternal, as well as apical images were obtained from all patients except from one, in whom parasternal view could not be collected due to the patient's inability to maintain the supine position. The primary endpoint was achieved in 23 patients (96%) for the apical view, in 20 patients (87%) for the parasternal view, and in all patients in at least one of the two views. The images' quality was maximal (i.e., score = 4) in at least one of the two windows in 19 patients (79%). Atrial fibrillation arrhythmia was the only clinical characteristics associated with a poor score outcome in both imaging windows (apical p = 0.022, parasternal p = 0.014). Conclusions These results provide the proof-of-concept for the feasibility of an automatic ultrasonographic image acquisition system associated with an AI algorithm for real-time monitoring of cardiac motion in patients with a history of VA.
【 授权许可】
CC BY
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202301300016190ZK.pdf | 801KB | download |