期刊论文详细信息
IEEE Photonics Journal
Reflective SFT-FBG Hybrid Micro-Probe for Simultaneous Measurement of Relative Humidity and Temperature
Yize Liu1  Zhankun Lv1  Panpan Niu1  Shuang Wang1  Tiegen Liu1  Junfeng Jiang1 
[1] School of Precision Instrument and Opto-electronics Engineering, Tianjin University, Tianjin, China;
关键词: Optical fiber probe;    S fiber taper;    Graphene oxide;    Fiber Bragg grating;    Relative humidity measurement;   
DOI  :  10.1109/JPHOT.2021.3136677
来源: DOAJ
【 摘 要 】

An optical fiber probe composed of a fiber Bragg grating (FBG) and a reflective S fiber taper (SFT) with graphene oxide (GO) film, for simultaneous relative humidity (RH) and temperature measurement, was proposed and experimentally demonstrated in this paper. The SFT is made into reflection type by depositing silver mirror at the fiber end face, which is highly sensitive to surrounding refractive index (RI). The GO nanosheets, as a two-dimensional hydrophilic nano material, is coated on the fiber surface, acting as a RH-to-RI converter. The simultaneous RH and temperature measurements are realized by monitoring the characteristic wavelength shifts of the SFT and FBG in the reflection spectrum. Experimental results show that the RH sensitivity of the probe is 161.1 pm/%RH during the range of 30%RH∼90%RH, and the temperature sensitivity is 14.1 pm/°C during the range of 10 °C∼50 °C. The proposed probe is compact, flexible, and easy to be fabricated, which has potential application for precise localized measurement in medical and biochemical fields.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次