Beta particles radiation doses have important applications in medicine. In particular, curved and symmetric as well as curved and asymmetric applicators containing the beta emitting 106Ru/106Rh isotopes are widely used in radiotherapy for the treatment of various ocular diseases. Nevertheless, a great problem in the use of these applicators is the inaccurate determination of the dose rates around them. Difficulties arise mainly because of the very short distances involved, and in this scenario theoretical calculation methods play an important role. In this work a simple approach based on the beta-point dose function integration over the total surface of each plaque was used to estimate the dose rates along their central axis. Results of relative dose rates for concave and symmetric (CCA, CCB, CXS, CCX/Y/Z, CCD, CGD and CCC) and concave and asymmetric (CIA, CIB/CIB-2, COB and COC) ruthenium/rhodium plaque types are shown.