期刊论文详细信息
Biosurface and Biotribology
The development of simulator testing for total knee replacements
Ioan Cracaoanu1  Casey Wise2  Amber Metcalfe2  Daniel D. Auger2  Chase Maag2 
[1] DePuy Synthes Leeds UK;DePuy Synthes Warsaw Indiana USA;
关键词: wear;    kinematics;    bone;    orthopaedics;    prosthetics;    biomechanics;   
DOI  :  10.1049/bsb2.12001
来源: DOAJ
【 摘 要 】

Abstract Professor Duncan Dowson was a pioneer in the field of tribology and simulator design. His work sparked many branches of research across orthopaedics. The first knee simulator described by Dowson was intended to measure the wear performance of early total knee replacements (TKRs). The industry has since advanced to achieve simulator designs with significant improvements including multi‐station, multi‐axis, multi‐control, and multi‐environmental capabilities. These simulators are used to test and compare not only wear, but also the kinematic/kinetic behaviour of TKRs and many other TKR design interactions prior to implantation. This has led to changes to the design of TKRs ranging from improvements to the tibial insert to the femoral component; all, in some way, thanks to Professor Duncan Dowson's inquisitive nature. This article provides a selective review to show the interdependencies of research and development endeavours starting with the evolution of knee simulators, the many advances in TKRs and finally the interconnection with cadaveric motion simulators.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次