GMS Hygiene and Infection Control | |
Microbiological evaluation of different reprocessing methods for cuffed and un-cuffed tracheostomy tubes in home-care and hospital setting | |
Leonhard, Matthias1  Zumtobel, Michaela1  Schneider-Stickler, Berit1  Assadian, Ojan2  | |
[1] Department of Otorhinolaryngology, Medical University of Vienna, Austria;Institute for Skin Integrity and Infection Prevention, University of Huddersfield, United Kingdom; | |
关键词: disinfection; dishwasher; ultrasound; sonification; cuff; reprocessing; tracheostomy tube; biofilm; infection control; medical device; | |
DOI : 10.3205/dgkh000262 | |
来源: DOAJ |
【 摘 要 】
Background: Manufacturers’ recommendations on cleaning of tracheostomy tubes focus on general warning information and non-specific manual cleaning procedures. The aim of this experimental study was to evaluate different reprocessing methods and to determine the mechanical integrity and functionality of tracheostomy tubes following reprocessing.Methods: Sixteen cuffed or un-cuffed tracheostomy tubes obtained from hospital in-patients were reprocessed using one of the following reprocessing methods: a) manual brushing and rinsing with tap water,brushing followed by disinfection with a glutaraldehyde solution, c) manual brushing followed machine-based cleaning in a dishwasher, and d) manual brushing followed by ultrasound cleaning in a commercially available ultrasound device. Microbial burden of the tubes before and after reprocessing was assessed by measurement of microbial colony-forming units per mL (CFU/mL) of rinsing fluid. After cleaning, tracheostomy tubes were investigated for loss of functionality. Findings: Manual brushing and rinsing with tap water reduced microbial colonization in average by 10 CFU/mL, but with poor reproducibility and reliability. Complete microbial reduction was achieved only with additional chemical or machine-based thermal disinfection. Ultrasound sonification yielded no further microbial reduction after manual brushing. Conclusion: Manual brushing alone will not result in complete eradication of microorganism colonising cuffed or un-cuffed tracheostomy tubes. However, manual cleaning followed by chemical or thermal disinfection may be regarded as safe and reproducible reprocessing method. If a machine-based reprocessing method is used for cuffed tubes, the cuffs’ ventilation hose must be secured in a safe position prior to thermal disinfection.
【 授权许可】
Unknown