期刊论文详细信息
Sensors
Fabrication of Parylene-Coated Microneedle Array Electrode for Wearable ECG Device
Sungbo Cho1  AfraizTariq Satti1  Hansang Kim2  Jangwoong Park2  Jinsoo Park3 
[1] Department of Electronics Engineering, Gachon University, Seongnam-si, Gyeonggi-do 13210, Korea;Department of Mechanical Engineering, Gachon University, Seongnam-si, Gyeonggi-do 13210, Korea;Gachon Advanced Institute for Health Science and Technology, Gachon University, Incheon 21999, Korea;
关键词: microneedle electrode;    ECG;    cardiovascular diseases;    parylene;    impedance;   
DOI  :  10.3390/s20185183
来源: DOAJ
【 摘 要 】

Microneedle array electrodes (MNE) showed immense potential for the sensitive monitoring of the bioelectric signals by penetrating the stratum corneum with high electrical impedance. In this paper, we introduce a rigid parylene coated microneedle electrode array and portable electrocardiography (ECG) circuit for monitoring of ECG reducing the motion artifacts. The developed MNE showed stability and durability for dynamic and long-term ECG monitoring in comparison to the typical silver-silver chloride (Ag/AgCl) wet electrodes. The microneedles showed no mechanical failure under the compression force up-to 16 N, but successful penetration of skin tissue with a low insertion force of 5 N. The electrical characteristics of the fabricated MNE were characterized by impedance spectroscopy with equivalent circuit model. The designed wearable wireless ECG monitoring device with MNE proved feasibility of the ECG recording which reduces the noise of movement artifacts during dynamic behaviors.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次