GMS Medizinische Informatik, Biometrie und Epidemiologie | |
Mapping of medical device data from ISO/IEEE 11073-10207 to HL7 FHIR | |
Ulrich, Hannes1  Ingenerf, Josef2  Andersen, Björn2  Riech, Kathrin Pia2  | |
[1] IT Center for Clinical Research, University of Lübeck, Germany;Institute of Medical Informatics, University of Lübeck, Germany; | |
关键词: iso/ieee 11073 sdc; hl7 fhir; medical communication standards; system interoperability; | |
DOI : 10.3205/mibe000222 | |
来源: DOAJ |
【 摘 要 】
The ISO/IEEE 11073-10207 Domain Information and Service Model focuses on cross-manufacturer device-to-device communication, whereas HL7 Fast Healthcare Interoperability Resources (FHIR) support the data exchange between software systems in health care. The contents of both standards overlap, yet they differ in structure. In this work, we present an Implementation Guide that contains a mapping of medical device data from ISO/IEEE 11073-10207 to HL7 FHIR in order to improve vendor-independent interoperability. This Implementation Guide helps to bridge the structural interoperability gap between these two communication architectures for medical devices and clinical information systems. This furthermore promotes the reuse of medical device data, e.g. for clinical research purposes.To facilitate the mapping while retaining contextual information despite the structural differences, it was only necessary to create five profiles for two FHIR resources. This Implementation Guide is described in relation to similar efforts including difficulties that arose during the implementation of the mapping. Approximately 25 problems occurred, but in all cases they are not critical and most of them could be solved for our use cases. In the future, this work will be merged with the Implementation Guide for point-of-care medical devices.
【 授权许可】
Unknown