EJNMMI Radiopharmacy and Chemistry | |
A suitable time point for quantifying the radiochemical purity of 225Ac-labeled radiopharmaceuticals | |
Patrick W. Causey1  Justin J. Wilson2  Alejandro Amor-Coarasa3  John W. Babich4  James M. Kelly4  Elizabeth Sweeney5  | |
[1] Canadian Nuclear Laboratories;Department of Chemistry and Chemical Biology, Cornell University;Department of Radiology, Albert Einstein College of Medicine;Department of Radiology, Molecular Imaging Innovations Institute (MI3), Weill Cornell Medicine;Division of Biostatistics, Department of Population Health Sciences, Weill Cornell Medicine; | |
关键词: Targeted alpha therapy; Ac-225; Radiopharmacy; Quality control; | |
DOI : 10.1186/s41181-021-00151-y | |
来源: DOAJ |
【 摘 要 】
Abstract Background As 225Ac-labeled radiopharmaceuticals continue to show promise as targeted alpha therapeutics, there is a growing need to standardize quality control (QC) testing procedures. The determination of radiochemical purity (RCP) is an essential QC test. A significant obstacle to RCP testing is the disruption of the secular equilibrium between actinium-225 and its daughter radionuclides during labeling and QC testing. In order to accelerate translation of actinium-225 targeted alpha therapy, we aimed to determine the earliest time point at which the RCP of an 225Ac-labeled radiopharmaceutical can be accurately quantified. Results Six ligands were conjugated to macrocyclic metal chelators and labeled with actinium-225 under conditions designed to generate diverse incorporation yields. RCP was determined by radio thin layer chromatography (radioTLC) followed by exposure of the TLC plate on a phosphor screen either 0.5, 2, 3.5, 5, 6.5, or 26 h after the plate was developed. The dataset was used to create models for predicting the true RCP for any pre-equilibrium measurement taken at an early time point. The 585 TLC measurements span RCP values of 1.8–99.5%. The statistical model created from these data predicted an independent data set with high accuracy. Predictions made at 0.5 h are more uncertain than predictions made at later time points. This is primarily due to the decay of bismuth-213. A measurement of RCP > 90% at 2 h predicts a true RCP > 97% and guarantees that RCP will exceed 90% after secular equilibrium is reached. These findings were independently validated using NaI(Tl) scintillation counting and high resolution gamma spectroscopy on a smaller set of samples with 10% ≤ RCP ≤ 100%. Conclusions RCP of 225Ac-labeled radiopharmaceuticals can be quantified with acceptable accuracy at least 2 h after radioTLC using various methods of quantifying particle emissions. This time point best balances the need to accurately quantify RCP with the need to safely release the batch as quickly as possible.
【 授权许可】
Unknown