期刊论文详细信息
Micromachines
Analytical, Numerical and Experimental Study of a Horizontal Electrothermal MEMS Microgripper for the Deformability Characterisation of Human Red Blood Cells
Marija Cauchi1  Pierluigi Mollicone1  Bertram Mallia2  Nicholas Sammut3  Ivan Grech3 
[1] Department of Mechanical Engineering, Faculty of Engineering, University of Malta, MSD 2080 Msida, Malta;Department of Metallurgy and Materials Engineering, Faculty of Engineering, University of Malta, MSD 2080 Msida, Malta;Department of Microelectronics and Nanoelectronics, Faculty of Information and Communications Technology, University of Malta, MSD 2080 Msida, Malta;
关键词: MEMS microgripper;    red blood cells;    electrothermal actuation;    analytical model;    finite element model;    fabrication;    PolyMUMPs™;    temperature;    displacement;    stresses;   
DOI  :  10.3390/mi9030108
来源: DOAJ
【 摘 要 】

Microgrippers are typical microelectromechanical systems (MEMS) that are widely used for micromanipulation and microassembly in both biological and micromanufacturing fields. This paper presents the design, modelling, fabrication and experimental testing of an electrothermal microgripper based on a ‘hot and cold arm’ actuator design that is suitable for the deformability characterisation of human red blood cells (RBCs). The analysis of the mechanical properties of human RBCs is of great interest in the field of medicine as pathological alterations in the deformability characteristics of RBCs have been linked to a number of diseases. The study of the microgripper’s steady-state performance is initially carried out by the development of a lumped analytical model, followed by a numerical model established in CoventorWare® (Coventor, Inc., Cary, NC, USA) using multiphysics finite element analysis. Both analytical and numerical models are based on an electothermomechanical analysis, and take into account the internal heat generation due to the applied potential, as well as conduction heat losses through both the anchor pads and the air gap to the substrate. The models are used to investigate key factors of the actuator’s performance including temperature distribution, deflection and stresses based on an elastic analysis of structures. Results show that analytical and numerical values for temperature and deflection are in good agreement. The analytical and computational models are then validated experimentally using a polysilicon microgripper fabricated by the standard surface micromachining process, PolyMUMPs™ (Durham, NC, USA). The microgripper’s actuation is characterised at atmospheric pressure by optical microscopy studies. Experimental results for the deflection of the microgripper arm tips are found to be in good agreement with the analytical and numerical results, with process-induced variations and the non-linear temperature dependence of the material properties accounting for the slight discrepancies observed. The microgripper is shown to actuate to a maximum opening displacement of 9μ m at an applied voltage of 3 V, thus being in line with the design requirement of an approximate opening of 8μ m for securing and characterising a RBC.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:7次