期刊论文详细信息
Nuclear Engineering and Technology
Investigation of Dose Distribution in Mixed Neutron-Gamma Field of Boron Neutron Capture Therapy using N-Isopropylacrylamide Gel
Yaser Kasesaz1  Hossein Khalafi1  Azim Khajeali2  Elham Bavarnegin3  Alireza Sadremomtaz3 
[1] Department of Physics, University of Guilan, P.O. Box 1841, Namjoo Street, Rasht, Iran;Medical Education Research Center, P.O. Box 5165665931, Golgasht Street, Tabriz, Iran;Nuclear Science and Technology Research Institute (NSTRI), P.O. Box 14395-836, North Kargar Street, Tehran, Iran;
关键词: BNCT;    Gel Dosimetry;    Mixed Field Dosimetry;    NIPAM Polymer Gel;    TRR BNCT Beam Line;   
DOI  :  10.1016/j.net.2016.07.012
来源: DOAJ
【 摘 要 】

Gel dosimeters have unique advantages in comparison with other dosimeters. Until now, these gels have been used in different radiotherapy techniques as a reliable dosimetric tool. Because dose distribution measurement is an important factor for appropriate treatment planning in different radiotherapy techniques, in this study, we evaluated the ability of the N-isopropylacrylamide (NIPAM) polymer gel to record the dose distribution resulting from the mixed neutron-gamma field of boron neutron capture therapy (BNCT). In this regard, a head phantom containing NIPAM gel was irradiated using the Tehran Research Reactor BNCT beam line, and then by a magnetic resonance scanner. Eventually, the R2 maps were obtained in different slices of the phantom by analyzing T2-weighted images. The results show that NIPAM gel has a suitable potential for recording three-dimensional dose distribution in mixed neutron-gamma field dosimetry.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:1次