Develop a high energy proton beam position monitor using linear contact image sensor
Huan Niu1  Tung-Yuan Hsiao1  Tzung-Yuang Chen1  Chien-Hsu Chen2 
[1] Accelerator Laboratory, Nuclear Science and Technology Development Center, National Tsing Hua University, Hsinchu, Taiwan;Corresponding author.;
关键词: Beam diagnostics;    Beam position monitor;    Beam profile monitor;    Scintillator;    Contact image sensor;    Ionization chamber;   
DOI  :  
来源: DOAJ
【 摘 要 】

A compact beam-position monitor was constructed using a linear contact image sensor attached to a plastic scintillator and tested using a 230 MeV proton beam. The results indicate that the beam position can be obtained in real-time, and the beam position with a precision of up to 0.03 mm. The compactness and high precision of the device hold considerable potential for it to be used as a beam-position monitor and offline, daily quality assurance monitor in hadron therapy. • The method can provide a high precision and high resolution beam position for flash irradiation in particle therapy in real-time. • The method using contact image sensor with scintillator does not require a long focal length for camera and it is free of image distortion. • The method can be integrated into medical particle accelerator for feedback control and daily quality assurance.