期刊论文详细信息
Applied Sciences
Demonstration of the Systematic Evaluation of an Optical Lattice Clock Using the Drift-Insensitive Self-Comparison Method
Yebing Wang1  Xiaotong Lu1  Hong Chang1  Chihua Zhou1  Benquan Lu1 
[1] Key Laboratory of Time and Frequency Primary Standards, National Time Service Center, Chinese Academy of Sciences, Xi’an 710600, China;
关键词: optical lattice clocks;    strontium atoms;    optical lattices;    laser cooling and trapping;   
DOI  :  10.3390/app11031206
来源: DOAJ
【 摘 要 】

The self-comparison method is a powerful tool in the uncertainty evaluation of optical lattice clocks, but any drifts will cause a frequency offset between the two compared clock loops and thus lead to incorrect measurement result. We propose a drift-insensitive self-comparison method to remove this frequency offset by adjusting the clock detection sequence. We also experimentally demonstrate the validity of this method in a one-dimensional 87Sr optical lattice clock. As the clock laser frequency drift exists, the measured frequency difference between two identical clock loops is (240 ± 34) mHz using the traditional self-comparison method, while it is (−15 ± 16) mHz using the drift-insensitive self-comparison method, indicating that this frequency offset is cancelled within current measurement precision. We further use the drift-insensitive self-comparison technique to measure the collisional shift and the second-order Zeeman shift of our clock and the results show that the fractional collisional shift and the second-order Zeeman shift are 4.54(28) × 10−16 and 5.06(3) × 10−17, respectively.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:1次