期刊论文详细信息
Journal of NeuroEngineering and Rehabilitation
Development of a battery-free ultrasonically powered functional electrical stimulator for movement restoration after paralyzing spinal cord injury
Serena Ng1  Monzurul Alam2  Shuai Li2  Yat Man Yam2  Yong-Ping Zheng2  Rakib Uddin Ahmed2  Suman Thakur3  Dan Tang4  Xiao-Yun Wang4 
[1] Community Rehabilitation Service Support Centre, Hospital Authority;Department of Biomedical Engineering, The Hong Kong Polytechnic University;Department of Chemical Sciences, Tezpur University;Guangdong Work Injury Rehabilitation Center;
关键词: Functional electrical stimulation;    Wireless power;    Ultrasound;    Piezoelectric;   
DOI  :  10.1186/s12984-019-0501-4
来源: DOAJ
【 摘 要 】

Abstract Background Functional electrical stimulation (FES) is used to restore movements in paretic limbs after severe paralyses resulting from neurological injuries such as spinal cord injury (SCI). Most chronic FES systems utilize an implantable electrical stimulator to deliver a small electric current to the targeted muscle or nerve to stimulate muscle contractions. These implanted stimulators are generally bulky, mainly due to the size of the batteries. Furthermore, these battery-powered stimulators are required to be explanted every few years for battery replacement which may result in surgical failures or infections. Hence, a wireless power transfer technique is desirable to power these implantable stimulators. Methods Conventional wireless power transduction faces significant challenges for safe and efficient energy transfer through the skin and deep into the body. Inductive and electromagnetic power transduction is generally used for very short distances and may also interfere with other medical measurements such as X-ray and MRI. To address these issues, we have developed a wireless, ultrasonically powered, implantable piezoelectric stimulator. The stimulator is encapsulated with biocompatible materials. Results The stimulator is capable of harvesting a maximum of 5.95 mW electric power at an 8-mm depth under the skin from an ultrasound beam with about 380 mW/cm2 of acoustic intensity. The stimulator was implanted in several paraplegic rats with SCI. Our implanted stimulator successfully induced several hindlimb muscle contractions and restored leg movement. Conclusions A battery-free miniature (10 mm diameter × 4 mm thickness) implantable stimulator, developed in the current study is capable of directly stimulating paretic muscles through external ultrasound signals. The required cost to develop the stimulator is relatively low as all the components are off the shelf.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次