| The effects of time-variance on impedance measurements:examples of a corroding electrode and a battery cell | |
| Nicolas Murer1  Jean-Paul Diard1  Bogdan Petrescu1  | |
| [1] Bio-Logic SAS, 4 Rue de Vaucanson, 38170 Seyssinet-Pariset; | |
| 关键词: EIS; non-stationarity; corrosion; in operando; battery cells; 4D impedance; | |
| DOI : 10.5599/jese.725 | |
| 来源: DOAJ | |
【 摘 要 】
When performing electrochemical impedance spectroscopy (EIS) measurements on a system, we must make sure it fulfills certain conditions. One of them is that it should be stationary that is to say, steady-state and time-invariant. Commonly studied systems are time-variant, for example a corroding electrode or a battery under operation. A corroding electrode sees its polarization resistance decrease with time. A passivating electrode sees its polarization resistance increase with time. These phenomena cause a deformation of the Nyquist impedance at low frequencies. This result was first simulated and validated by experimental measurements on a corroding steel sample undergoing uniform corrosion. The effect of performing impedance measurements on a discharging battery was also shown. Several methods are available to check and correct time-variance. The non-stationary distortion (NSD) indicator is used to separate valid and invalid data samples and the so called “4D impedance” method can easily produce instantaneous impedance data.