Comparison between ultrasound-guided TIVAD via the right innominate vein and the right internal jugular vein approach | |
Yong Jin1  Xingwei Sun1  Xuming Bai1  Jiaofeng Shen2  Zhixiang Zhuang2  Ziyang Yu3  | |
[1] Department of Intervention, The Second Affiliated Hospital of Soochow University;Department of Oncology, The Second Affiliated Hospital of Soochow University;Department of Ultrasound, The Second Affiliated Hospital of Soochow University; | |
关键词: Totally implantable venous access device; Innominate vein; Internal jugular vein; US-guided; Cancer; | |
DOI : 10.1186/s12893-019-0651-0 | |
来源: DOAJ |
【 摘 要 】
Abstract Background To compare the efficacy and safety of right internal jugular vein (IJV) approach and right innominate vein (INV) approach for US-guided totally implantable venous access devices (TIVADs), and to explore the advantages and disadvantages of the two approaches. Methods Six hundred and nineteen adult patients had long-term infusion and chemotherapy needs and inconvenience of peripheral venous infusion. Right INV approach was used to implant 339 cases of TIVADs, and right IJV approach was used to implant 280 cases of TIVADs. The success rate of one-time catheterization and the incidence of complications in the two groups were retrospectively analyzed. Results All patients were successfully implanted in TIVAD. The success rates of one-time puncture in INV group and IJV approach group were 98.53% (334/339) and 95.36% (267/280), respectively. There was significant difference between the two groups (P = 0.020). The incidence of perioperative complications and long-term complications in the right INV group were 1.18% (4/339) and 3.54% (12/339), respectively, while those in the right IJV group were 1.43% (4280) and 3.93% (11280). There was no significant difference in the incidence of perioperative or long-term complications between the two groups (P = 0.785, P = 0.799, respectively). Conclusions US-guided TIVADs via the right INV approach and the right IJV approach are both safe and reliable. The right INV approach improves the one-time puncture success rate, as long as the technique is properly operated, serious complications rarely occur.
【 授权许可】
Unknown